FDA Drug recall · Reported February 13, 2013
EstroGel (estradiol gel) 0.06%, contained in 50-gr metered pumps, 1.25g/pump depression, for use as physician starter samples; product is labeled in part ***Manufactured for ASCEND Therapeutics Herndon, VA 20170 by DPT Laboratories San Antonio TX 78215***
Defective Container: Pump head detaching from the canister unit upon removal of the overcap.
Ascend Therapeutics Inc recalled EstroGel (estradiol gel) 0.06%, contained in 50-gr metered pumps, 1.25g/pump depression, … - a low-severity action.
EstroGel (estradiol gel) 0.06%, contained in 50-gr metered pumps, 1.25g/pump depression, … was recalled by Ascend Therapeutics Inc in February 13, 2013. Reason: Defective Container: Pump head detaching from the canister unit upon removal of the overcap.. Check the official notice for the remedy. Verify recall #D-146-2013 with the FDA Drug before acting.
The recall
Ascend Therapeutics Inc issued this low-severity FDA Drug recall-Defective Container: Pump head detaching from the canister unit upon removal of the overcap..
- Low
- severity level
- Class III
- classification
- February 13, 2013
- reported
Sourced from official FDA Drug enforcement records. Verify recall #D-146-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified low severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
EstroGel (estradiol gel) 0.06%, contained in 50-gr metered pumps, 1.25g/pump depression, for use as physician starter samples; product is labeled in part ***Manufactured for ASCEND Therapeutics Herndon, VA 20170 by DPT Laboratories San Antonio TX 78215***
Reason for recall
Defective Container: Pump head detaching from the canister unit upon removal of the overcap.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Low (Class III) |
| Status | Terminated |
| Recall number | D-146-2013 |
| Date reported | February 13, 2013 |
| Date initiated | November 19, 2012 |
| Recalling firm | Ascend Therapeutics Inc |
| Firm location | Herndon, VA |
| Affected scope | 4, 835 bottles |
| Distribution | Products were distributed nationwide. |
Profile values are sourced directly from the official FDA Drug enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (D-146-2013) and product description against the item you own. Search the archive
- Confirm the current status with FDA Drug before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Drug record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Low), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026