Allura Xper FD20; Model Numbers: (1) 722006, (2) 722012, (3) 722028; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Reported: October 8, 2025 Initiated: September 3, 2025 #Z-0005-2026
Product Description
Allura Xper FD20; Model Numbers: (1) 722006, (2) 722012, (3) 722028; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;
Reason for Recall
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Details
- Recalling Firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Units Affected
- 5067 units (1626 US, 3441 OUS)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Curacao, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Finland, France, Georgia, Germany, Greece, Guinea, Honduras, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Malta, Mauritius, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, T¿rkiye, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Utd.Arab.Emir., Uzbekistan, Venezuela, Vietnam, Yemen.
- Location
- Best, N/A
Frequently Asked Questions
What product was recalled? ▼
Allura Xper FD20; Model Numbers: (1) 722006, (2) 722012, (3) 722028; Associated 510(k)s: K102005, K130638, K130842, K133292, K141979, K162859, K161563;. Recalled by PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.. Units affected: 5067 units (1626 US, 3441 OUS).
Why was this product recalled? ▼
The BIOS battery may deplete faster than initially anticipated during the design phase. When the battery is depleted, the start-up process of the system will be halted, and the system will not start. No user messages are shown before the battery is low on power or depleted.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on October 8, 2025. Severity: Moderate. Recall number: Z-0005-2026.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11