Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications during primary and revision Total Knee Arthroplasty (TKA) when the Triathlon Knee System is selected for use. Catalog Number: 5521-B-200

Reported: October 16, 2024 Initiated: September 4, 2024 #Z-0023-2025 16 units units

Howmedica Osteonics Corp. issued this FDA Devices recall on October 16, 2024. Classified as Moderate severity (Class II). Approximately 16 units units are affected. The recall was issued because: Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of eit…. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0023-2025) was formally reported on October 16, 2024, with the manufacturer initiating the action on September 4, 2024. It is classified under Moderate severity (Class II), with a current status of Ongoing. Howmedica Osteonics Corp. is listed as the recalling firm, operating out of Mahwah, NJ. Federal records indicate 16 units units are affected.

The documented reason for this recall is: Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert. Distribution data in the federal record shows the product reached: US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI. Stryker issued Urgent Medical Device Recall (UMDR) PFA 3554908 by third party Sedgwick on September 04, 2024, via UPS 2-day air. Letter state…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 2 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

16 units

Related Recalls

6

6 from same agency

Product Description

Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications during primary and revision Total Knee Arthroplasty (TKA) when the Triathlon Knee System is selected for use. Catalog Number: 5521-B-200

Reason for Recall

Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.

Details

Recalling Firm
Howmedica Osteonics Corp.
Units Affected
16 units
Distribution
US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI. Stryker issued Urgent Medical Device Recall (UMDR) PFA 3554908 by third party Sedgwick on September 04, 2024, via UPS 2-day air. Letter states reason for recall, health risk and action to take: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware or organizations who have consigned product. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list (Table 1). 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form: " Medical Facility: Please sign the Business Reply Form and email to: strykerortho8441@sedgwick.com/ Fax: (800) 871-7417 " Stryker Branch/Agency: please sign the AdobeSign form arriving to you via email. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return the devices to Stryker using the following address: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 Ref. PFA 3554908 Please add the tracking number Please assist us in meeting our regulatory obligation by completing and emailing the attached Urgent Medical Device Recall Business Reply Form within 5 days. A response is required, even though you may not have any physical inventory on site. If you have any questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com
Location
Mahwah, NJ

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0023-2025
Date reported October 16, 2024
Date initiated September 4, 2024
Recalling firm Howmedica Osteonics Corp.
Units affected 16 units
Distribution US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI. Stryker issued Urgent Medical Device Recall (UMDR) PFA 3554908 by third party Sedgwick on September 04, 2024, via UPS 2-day air. Letter states reason for recall,…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

16 units units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications during primary and revision Total Knee Arthroplasty (TKA) when the Triathlon Knee System is selected for use. Catalog Number: 5521-B-200. Recalled by Howmedica Osteonics Corp.. Units affected: 16 units.
Why was this product recalled?
Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 16, 2024. Severity: Moderate. Recall number: Z-0023-2025.
Where was the recalled product distributed?
Distribution: US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI. Stryker issued Urgent Medical Device Recall (UMDR) PFA 3554908 by third party Sedgwick on September 04, 2024, via UPS 2-day air. Letter states reason for recall, health risk and action to take: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware or organizations who have consigned product. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list (Table 1). 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form: " Medical Facility: Please sign the Business Reply Form and email to: strykerortho8441@sedgwick.com/ Fax: (800) 871-7417 " Stryker Branch/Agency: please sign the AdobeSign form arriving to you via email. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return the devices to Stryker using the following address: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 Ref. PFA 3554908 Please add the tracking number Please assist us in meeting our regulatory obligation by completing and emailing the attached Urgent Medical Device Recall Business Reply Form within 5 days. A response is required, even though you may not have any physical inventory on site. If you have any questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email SO_M_PRODUCT_FIELD_ACTION_RESPONSE@stryker.com.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0023-2025) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).