DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer

Reported: October 8, 2025 Initiated: August 4, 2025 #Z-0032-2026

Product Description

DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer

Reason for Recall

On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.

Details

Recalling Firm
Beckman Coulter Inc.
Units Affected
116
Distribution
Internationally distribution to the countries of Poland, India, Dominican Republic, Lebanon, Italy, Brazil, El Salvador, Saudi Arabia, Germany, France, Nepal, Canada, Chile, Spain, Malaysia, Switzerland, Morocco, Philippines, Curacao, Qatar, United Arab Emirates, Hungary, Czechia, Oman, New Zealand, Thailand, Indonesia, Algeria, United Kingdom of Great Britain and Northern Ireland, Namibia, Turkey, Republic of Korea
Location
Brea, CA

Frequently Asked Questions

What product was recalled?
DxC 500 AU Module w/ISE, DxC 500i, With ISE, REF: C63522 and DxC 500 AU Module, DxC 500i, Without ISE, REF: C6352; part of the DxC 500i Clinical Analyzer. Recalled by Beckman Coulter Inc.. Units affected: 116.
Why was this product recalled?
On a clinical analyzer, when ordering a Clinical Chemistry (CC) combination test simultaneously with an Immunoassay (IA) test, if the IA test is processed between the CC tests, CC sample status will remain In Progress with no errors, may cause processing delays.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 8, 2025. Severity: Moderate. Recall number: Z-0032-2026.