TetraCXP Software System, Part Number A40051. Allows simultaneous identification and enumeration of total CD3+, total CD4+, total CD8+, dual CD3+/CD4+, dual CD3+/CD8+ and/or total CD3+, CD19+ and CD3+/CD56+ lymphocyte percentages and absolute counts in whole blood by flow cytometry. The systems also provide the CD4/CD8 ratio when using CD45-FITC/ CD4-RD1/CD8-ECD/CD3-PC5, and total lymphocyte percentage when using CD45-FITC/CD56-RD1/ CD19-ECD/CD3-PC5.
Product Description
TetraCXP Software System, Part Number A40051. Allows simultaneous identification and enumeration of total CD3+, total CD4+, total CD8+, dual CD3+/CD4+, dual CD3+/CD8+ and/or total CD3+, CD19+ and CD3+/CD56+ lymphocyte percentages and absolute counts in whole blood by flow cytometry. The systems also provide the CD4/CD8 ratio when using CD45-FITC/ CD4-RD1/CD8-ECD/CD3-PC5, and total lymphocyte percentage when using CD45-FITC/CD56-RD1/ CD19-ECD/CD3-PC5.
Reason for Recall
The recall was initiated because Beckman Coulter has determined that the specimen and prepared sample stability claims stated in the product labeling for tetraCXP software and CYTO-STAT reagents (CD45-FITC/CD4-RD1/CD8-ECD/CD3-PC5 and CD45-FITC/CD56-RD1/CD19-ECD/CD3-PC5) need to be updated for alignment with current standards and clinical decision-making associated with lymphocyte immunophenotyping
Details
- Recalling Firm
- Beckman Coulter Inc.
- Units Affected
- 146 units total (37 units in US)
- Distribution
- Worldwide Distribution - USA (nationwide) and the following countries: Albania Australia Austria Bahrain Bangladesh Barbados Belarus Belgium Bosnia and Herzegovina Bulgaria Burundi Canada Chile China Colombia Croatia Czech Republic Denmark Ecuador El Salvador Finland France French Guiana French Polynesia Germany Greece Guadeloupe Hong Kong India Indonesia Iraq Ireland Israel Italy Japan Kenya Korea, Republic of Kuwait Libyan Arab Jamahiriya Lithuania Macao Malaysia Mexico Monaco Morocco Namibia Netherlands New Zealand Nigeria Norway Oman Pakistan Panama Poland Puerto Rico Qatar Reunion Romania Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sudan Sweden Switzerland Taiwan Turkey United Arab Emirates United Kingdom Uruguay
- Location
- Brea, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11