FDA Devices recall · Reported October 16, 2019
Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers pr
GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris ar…
Ohmeda Medical recalled Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and … - a moderate-severity action.
Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and … was recalled by Ohmeda Medical in October 16, 2019. Reason: GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned…. Check the official notice for the remedy. Verify recall #Z-0041-2020 with the FDA Devices before acting.
The recall
Ohmeda Medical issued this moderate-severity FDA Devices recall-GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned….
- Moderate
- severity level
- Class II
- classification
- October 16, 2019
- reported
Sourced from official FDA Devices enforcement records. Verify recall #Z-0041-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
Recall Insight
This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Medical Devices recalls over time
Where this recall sits in its category - 41,021 medical devices recalls on record
Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Product description
Giraffe Omnibed - Product Usage: The OmniBed is a combination of an infant incubator and an infant warmer. The device can be operated as an incubator or as a warmer and can transition from one mode to the other on user s demand. It cannot be operated in both modes at the same time. Incubators and warmers provide heat in a controlled manner to neonates who are unable to thermo-regulate based on their own physiology. Incubators provide an enclosed, temperature-controlled environment and warmers provide infrared heat in an open environment. They may also be used for short periods of time to facilitate the neonate s transition from the uterus to the external environment. This device may incorporate a Servo Controlled Oxygen Delivery System. This is indicated to provide stable oxygen concentration within the infant compartment at the value set by the operator (21-65%).
Reason for recall
GE Healthcare has become aware of a potential safety issue that can occur if the canopy seals are not cleaned appropriately. Healthcare professionals reported buildup of debris around the edge of the canopy seal which can lead to a source of infection. Upon review, the Agency concluded that the cleaning and care guidelines provided by the firm were inadequate for effective cleaning of the device. In particular, the instructions do not instruct for removal of canopy seal prior to cleaning, which leads to the buildup of debris.
Recall Profile
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0041-2020 |
| Date reported | October 16, 2019 |
| Date initiated | October 5, 2018 |
| Recalling firm | Ohmeda Medical |
| Firm location | Laurel, MD |
| Affected scope | N/A |
| Distribution | Worldwide distribution- Domestic nationwide and foreign distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
- Check the recall number (Z-0041-2020) and product description against the item you own. Search the archive
- Confirm the current status with FDA Devices before acting, recall details can be updated.
- Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Where was the recalled product distributed? ▼
Should I stop using a recalled medication or medical device? ▼
Nationwide recalls with similar severity or category timing
Drug and device notices outside this firm and state, matched by severity timing and product category.
Similar severity · nearest report date
Same archive severity grade (Moderate), closest by report date among other firms and states.
Similar Medical Devices · nearest report date
Same product category, closest by report date among other firms and states (deduped against the severity set).
Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls
Methodology · Corrections · Archive as of August 2026