Belmont Rapid Infuser, a Fluid Management System, Model FMS2000.

Reported: October 24, 2012 Initiated: September 25, 2012 #Z-0058-2013

Product Description

Belmont Rapid Infuser, a Fluid Management System, Model FMS2000.

Reason for Recall

Leak from a saline bag caused saline to drip along the back of the Rapid infuser close to the Power Entry Module causing an electrical flash at the power entry area.

Details

Units Affected
2,913 units
Distribution
Worldwide Distribution including Nationwide (USA) and the countries of Australia, Austria, Belgium, Brazil, Canada, China, Denmark, England, Estonia, France, Germany, Hong Kong, India, Israel, Italy, Jordan, Mongolia, Netherlands, New Zealand, Norway, Pakistan, Peru, Puerto Rico, S. Korea, Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, UK, United Arab Emirates, and Vietnam.
Location
Billerica, MA

Frequently Asked Questions

What product was recalled?
Belmont Rapid Infuser, a Fluid Management System, Model FMS2000.. Recalled by Belmont Instrument Corporation. Units affected: 2,913 units.
Why was this product recalled?
Leak from a saline bag caused saline to drip along the back of the Rapid infuser close to the Power Entry Module causing an electrical flash at the power entry area.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 24, 2012. Severity: Moderate. Recall number: Z-0058-2013.