PlainRecalls
FDA Devices Moderate Class II Ongoing

CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with heat exchange devices.

Reported: October 13, 2021 Initiated: June 24, 2021 #Z-0069-2022

Product Description

CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with heat exchange devices.

Reason for Recall

There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potentially aerosolized into the operating room during surgery which could result in patient infection.

Details

Recalling Firm
CardioQuip, LLC
Units Affected
1380 units
Distribution
Worldwide distribution - US Nationwide distribution including in the states of TX, PA, FL, CA, HI, SC, OH, AK, NY, IL, AR, WI, MI, AL, KS, NJ, GA, AZ, KY, MS, TN, LA, DE, MA, VA, CO, NE, NC, NH, ME, WV, PA, MO, MN, IN, CT, NH, NV, UT and the country of Venezuela.
Location
College Station, TX

Frequently Asked Questions

What product was recalled?
CardioQuip, Modular Cooler-Heater 1000(i), Refrigeration Module, Model: MCH-10RMS, UDI: (01)00860000846127 used with heat exchange devices.. Recalled by CardioQuip, LLC. Units affected: 1380 units.
Why was this product recalled?
There is the potential bacterial including Nontuberculous mycobacterial (NTM) to grow in the water system of heater-cooler devices which could result in contaminated water potentially aerosolized into the operating room during surgery which could result in patient infection.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 13, 2021. Severity: Moderate. Recall number: Z-0069-2022.