Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Reported: October 19, 2022 Initiated: May 30, 2022 #Z-0087-2023

Product Description

Positron Emission Tomography and Computed Tomography System, Model: uMI 550

Reason for Recall

The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.

Details

Units Affected
32
Distribution
Distribution to US states of TX, WI, CA, and MI
Location
Shanghai, N/A

Frequently Asked Questions

What product was recalled?
Positron Emission Tomography and Computed Tomography System, Model: uMI 550. Recalled by Shanghai United Imaging Healthcare Co., Ltd.. Units affected: 32.
Why was this product recalled?
The wireless VSM module of a mobile PET/CT system, operating in an environment with strong Wi-Fi signals, may experience ECG signal and respiratory signal loss due to Wi-Fi interference. ECG and respiratory signal loss during acquisition can result in the failure of ECG and respiratory-gated reconstruction of the PET scan, which may require rescanning of the patient.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 19, 2022. Severity: Moderate. Recall number: Z-0087-2023.