PlainRecalls
FDA Devices Moderate Class II Ongoing

N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299

Reported: October 26, 2022 Initiated: August 30, 2022 #Z-0094-2023

Product Description

N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299

Reason for Recall

N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).

Details

Units Affected
10,201 packs
Distribution
Nationwide
Location
Tarrytown, N/A

Frequently Asked Questions

What product was recalled?
N Antiserum to Human IgG, 5mL Variant for BN II System and BN ProSpec System- IVD Siemens Material Number (SMN): 10446299. Recalled by SIEMENS HEALTHCARE DIAGNOSTICS INC. Units affected: 10,201 packs.
Why was this product recalled?
N Antiserum to Human IgG are not meeting the current High-Dose Hook Effect expectation of up to 648 mg/L for IgGU (urine) and up to 2290 mg/L for IgGC (CSF).
Which agency issued this recall?
This recall was issued by the FDA Devices on October 26, 2022. Severity: Moderate. Recall number: Z-0094-2023.