GE Healthcare CARESCAPE Monitor B450 Product Usage: Intended uses of CARESCAPE B450 in 510(k) countries [UR_CRSCP-885]: The CARESCAPE Monitor B450 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CARESCAPE Monitor B450 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B450 is indicated for monitoring of: -hemodynamic (including ECG
Product Description
GE Healthcare CARESCAPE Monitor B450 Product Usage: Intended uses of CARESCAPE B450 in 510(k) countries [UR_CRSCP-885]: The CARESCAPE Monitor B450 is a multi-parameter patient monitor intended for use in multiple areas and intrahospital transport within a professional healthcare facility. The CARESCAPE Monitor B450 is intended for use on adult, pediatric, and neonatal patients and on one patient at a time. The CARESCAPE Monitor B450 is indicated for monitoring of: -hemodynamic (including ECG, ST segment, arrhythmia detection, ECG diagnostic analysis and measurement, invasive pressure, non-invasive blood pressure, pulse oximetry, cardiac output (thermodilution and pulse contour), temperature, mixed venous oxygen saturation, and central venous oxygen saturation, -Respiratory (impedance respiration, airway gases (CO2, O2, N2O and anesthetic agents), and spirometry) -neurophysiological status (including electroencephalography, Entropy, Bispectral Index (BIS), and neuromuscular transmission).
Reason for Recall
When multiple CARESCAPEMonitor B450 or B850 units are connected to the same network and a network overload occurs for a prolonged time, the monitors may simultaneously restart as designed. The monitor restart will not be completed until the network issue has been corrected.
Details
- Recalling Firm
- GE Healthcare Finland Oy
- Units Affected
- 59,000 units
- Distribution
- Worldwide Distribution-US Nationwide
- Location
- Helsinki
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11