Severity
Critical
FDA Devices recall · Reported October 27, 2021
Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line ev…
Smiths Medical Asd Inc. recalled CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-… - a critical-severity action.
CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-… was recalled by Smiths Medical Asd Inc. in October 27, 2021. Reason: Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD d…. Check the official notice for the remedy. Verify recall #Z-0107-2022 with the FDA Devices before acting.
The recall
Smiths Medical Asd Inc. issued this critical-severity FDA Devices recall-Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD d….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0107-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0107-2022) was formally reported on October 27, 2021, with the manufacturer initiating the action on March 24, 2020. It is classified under Critical severity (Class I), with a current status of Terminated. Smiths Medical Asd Inc. is listed as the recalling firm, operating out of Minneapolis, MN. Federal records list the affected scope as 642 units.
The documented reason for this recall is: Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump an… Distribution data in the federal record shows the product reached: Distribution to European countries: BE, BH, CH, DE, DK, ES, FI, FR, GB, IT, HL, NO, PT, SE. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,861 medical devices recalls on record
Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified high severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Critical
Affected scope
642 units
Related Recalls
6
6 from same agency
CADD-Solis VIP ambulatory infusion pump, Model Numbers: a. 21-2120-0105-02L b. 21-2120-0105-03L c. 21-2120-0105-06L d. 21-2120-0105-07L e. 21-2120-0105-08L f. 21-2120-0105-12L g. 21-2120-0105-13L h. 21-2120-0105-14L i. 21-2120-0105-15L j. 21-2120-0105-17L k. 21-2120-0105-50L l. 21-2127-0105-02L m. 21-2127-0105-06L n. 21-2127-0105-50L
Specific Pumps may exhibit intermittent performance in the AILD (Air in Line Detector) function. If an AILD does not sufficiently discern fluid from air in line, an air in line event may not be recognized by the pump and may not alarm to notify the clinician.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Critical (Class I) |
| Status | Terminated |
| Recall number | Z-0107-2022 |
| Date reported | October 27, 2021 |
| Date initiated | March 24, 2020 |
| Recalling firm | Smiths Medical Asd Inc. |
| Firm location | Minneapolis, MN |
| Affected scope | 642 units |
| Distribution | Distribution to European countries: BE, BH, CH, DE, DK, ES, FI, FR, GB, IT, HL, NO, PT, SE |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Two federal-filing-derived peer sets for this notice, both outside the same recalling firm and filing state so the neighborhoods are not firm or geographic containment.
Same archive severity grade (Critical), closest by report date among other firms and states.
Same product category, closest by report date among other firms and states (deduped against the severity set).
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
Showing 5 of 40909 other recalls in this category.
Boston Scientific Corporation · 2026-07-29
Boston Scientific Corporation · 2026-07-29
AVID Medical, Inc. · 2026-07-29
Penumbra, Inc. · 2026-07-29
Alphatec Spine, Inc. · 2026-07-29
Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported October 27, 2021. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of August 2026.