Vanguard Distal Femoral Augment 75 X 5 RL/LM Product Usage:  Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.  Correction of varus, valgus, or posttraumatic deformity.  Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.  The device is a single use implant intended for implantation with bone cement.

Reported: October 26, 2016 Initiated: August 9, 2016 #Z-0111-2017

Product Description

Vanguard Distal Femoral Augment 75 X 5 RL/LM Product Usage:  Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.  Correction of varus, valgus, or posttraumatic deformity.  Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.  The device is a single use implant intended for implantation with bone cement.

Reason for Recall

Contents of the package do not match the product labeling. Product is labeled: Vanguard Distal Femoral Augment 75 X 5 RL/LM Carton contains: Vanguard Distal Femoral Augment 75 X 5 LL/RM Risks: Adverse health outcomes that could occur as a result of this event are a delay less than 30 minutes and a different size augment being used to complete the procedure. As the augment will not assemble with femoral component due to its shape, it would be easily identified during surgery.

Details

Recalling Firm
Zimmer Biomet, Inc.
Units Affected
5
Distribution
Worldwide Distribution - US Nationwide in the states of NJ, MI, FL and the countries of Sweden and Canada.
Location
Warsaw, IN

Frequently Asked Questions

What product was recalled?
Vanguard Distal Femoral Augment 75 X 5 RL/LM Product Usage:  Painful and disabled knee joint resulting from osteoarthritis, rheumatoid arthritis, traumatic arthritis where one or more compartments are involved.  Correction of varus, valgus, or posttraumatic deformity.  Correction or revision of unsuccessful osteotomy, arthrodesis, or failure of previous joint replacement procedure.  The device is a single use implant intended for implantation with bone cement.. Recalled by Zimmer Biomet, Inc.. Units affected: 5.
Why was this product recalled?
Contents of the package do not match the product labeling. Product is labeled: Vanguard Distal Femoral Augment 75 X 5 RL/LM Carton contains: Vanguard Distal Femoral Augment 75 X 5 LL/RM Risks: Adverse health outcomes that could occur as a result of this event are a delay less than 30 minutes and a different size augment being used to complete the procedure. As the augment will not assemble with femoral component due to its shape, it would be easily identified during surgery.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 26, 2016. Severity: Moderate. Recall number: Z-0111-2017.