Severity
Moderate
FDA Devices recall · Reported October 28, 2020
The CHF-CB30S Device did not have a 510(k) clearance. Additionally, the device has been associated with breakage of the insertion section and protrusion of metal parts, as well as…
Olympus Corporation of the Americas recalled Olympus CHF-CB30S Choledoscope Intended Use: The device was designed to be used with … - a moderate-severity action.
Olympus CHF-CB30S Choledoscope Intended Use: The device was designed to be used with … was recalled by Olympus Corporation of the Americas in October 28, 2020. Reason: The CHF-CB30S Device did not have a 510(k) clearance. Additionally, the device has been associated with break…. Check the official notice for the remedy. Verify recall #Z-0119-2021 with the FDA Devices before acting.
The recall
Olympus Corporation of the Americas issued this moderate-severity FDA Devices recall-The CHF-CB30S Device did not have a 510(k) clearance. Additionally, the device has been associated with break….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0119-2021 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0119-2021) was formally reported on October 28, 2020, with the manufacturer initiating the action on August 31, 2020. It is classified under Moderate severity (Class II), with a current status of Terminated. Olympus Corporation of the Americas is listed as the recalling firm, operating out of Center Valley, PA. Federal records list the affected scope as 193 devices distributed in the U.S. 548 devices distributed globally..
The documented reason for this recall is: The CHF-CB30S Device did not have a 510(k) clearance. Additionally, the device has been associated with breakage of the insertion section and protrusion of metal parts, as well as breakage and displacement of the rubber… Distribution data in the federal record shows the product reached: US Nationwide Distribution in states of: AK, AZ, CA, FL, GA, IL, LA, MA, ME, MI, MN, MS, NC, NM, NJ, NY, OR, SC, TN, WA, and WI; and International to: Japan.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
193 devices distributed in the U.S. 548 devices distributed globally.
Related Recalls
6
3 from same agency
Olympus CHF-CB30S Choledoscope Intended Use: The device was designed to be used with an Olympus TV camera system, light source, TV monitor, endo-therapy accessories and other ancillary equipment for endoscopic diagnosis and treatment within the biliary tract (common bile duct, cystic duct, and hepatic duct).
The CHF-CB30S Device did not have a 510(k) clearance. Additionally, the device has been associated with breakage of the insertion section and protrusion of metal parts, as well as breakage and displacement of the rubber on the bending section during surgical procedures.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0119-2021 |
| Date reported | October 28, 2020 |
| Date initiated | August 31, 2020 |
| Recalling firm | Olympus Corporation of the Americas |
| Firm location | Center Valley, PA |
| Affected scope | 193 devices distributed in the U.S. 548 devices distributed globally. |
| Distribution | US Nationwide Distribution in states of: AK, AZ, CA, FL, GA, IL, LA, MA, ME, MI, MN, MS, NC, NM, NJ, NY, OR, SC, TN, WA, and WI; and International to: Japan. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported October 28, 2020. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.