Severity
Moderate
FDA Devices recall · Reported October 31, 2012
Ortho sensor in Sunrise, FL is recalling the Graphics User Interface Software 1.8 (which was updated to version 1.82, validated on 02/13/12). Recall was initiated due to customer …
Orthosensor recalled "****ORTHO SENSOR**12-123-A Knee Trial for Trithion***1560 Saw grass Corporate Pkwy***4T… - a moderate-severity action.
"****ORTHO SENSOR**12-123-A Knee Trial for Trithion***1560 Saw grass Corporate Pkwy***4T… was recalled by Orthosensor in October 31, 2012. Reason: Ortho sensor in Sunrise, FL is recalling the Graphics User Interface Software 1.8 (which was updated to versi…. Check the official notice for the remedy. Verify recall #Z-0122-2013 with the FDA Devices before acting.
The recall
Orthosensor issued this moderate-severity FDA Devices recall-Ortho sensor in Sunrise, FL is recalling the Graphics User Interface Software 1.8 (which was updated to versi….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0122-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0122-2013) was formally reported on October 31, 2012, with the manufacturer initiating the action on February 4, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Orthosensor is listed as the recalling firm, operating out of Sunrise, FL. Federal records list the affected scope as 9.
The documented reason for this recall is: Ortho sensor in Sunrise, FL is recalling the Graphics User Interface Software 1.8 (which was updated to version 1.82, validated on 02/13/12). Recall was initiated due to customer feedback which included the line of site… Distribution data in the federal record shows the product reached: Nationwide Distribution including the states of CA, FL, MI and NY.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
9
Related Recalls
6
3 from same agency
"****ORTHO SENSOR**12-123-A Knee Trial for Trithion***1560 Saw grass Corporate Pkwy***4Th floor***Sunrise, FL 33323***www.orthosensor.com***" Rx Only*** Sterile/EO***Made exclusively for Stryker Orthopaedics***. Is the graphic user interface software associated with the graphic user interface, which is loaded onto the linkstations.
Ortho sensor in Sunrise, FL is recalling the Graphics User Interface Software 1.8 (which was updated to version 1.82, validated on 02/13/12). Recall was initiated due to customer feedback which included the line of site between the knee ba lancer and the Link station was obstructed by OR metal stands or trays or the knee balancer device was placed atop of or near the metal stands or trays, interru
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0122-2013 |
| Date reported | October 31, 2012 |
| Date initiated | February 4, 2012 |
| Recalling firm | Orthosensor |
| Firm location | Sunrise, FL |
| Affected scope | 9 |
| Distribution | Nationwide Distribution including the states of CA, FL, MI and NY. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported October 31, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.