FDA Devices Moderate Class II Terminated

Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that passes through and works in conjunction with an introducer cannula to remove intervertebral disc nucleus pulposus material.

Reported: November 13, 2013 Initiated: November 4, 2013 #Z-0127-2014

Product Description

Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that passes through and works in conjunction with an introducer cannula to remove intervertebral disc nucleus pulposus material.

Reason for Recall

Stryker has identified a potential for the auger and/or cannula to fracture when the Dekompressor is used contrary to the warnings in the Instructions For Use (IFU). Specifically, manually bending the cannula can weaken the device and may result in a fracture of the auger and/or cannula. This could result in the need for additional medical intervention to remove the broken piece, pain or loss of m

Details

Units Affected
20,385 total distributed since 2004
Distribution
Nationwide Distribution including the states of AR, AZ, CA, CO, FL, GA, IL, KS, MS, ND, and NE.
Location
Portage, MI

Frequently Asked Questions

What product was recalled?
Dekompressor Percutaneous Discectomy Probe. The Dekompressor is a single use disposable discectomy probe that passes through and works in conjunction with an introducer cannula to remove intervertebral disc nucleus pulposus material.. Recalled by Stryker Instruments Div. of Stryker Corporation. Units affected: 20,385 total distributed since 2004.
Why was this product recalled?
Stryker has identified a potential for the auger and/or cannula to fracture when the Dekompressor is used contrary to the warnings in the Instructions For Use (IFU). Specifically, manually bending the cannula can weaken the device and may result in a fracture of the auger and/or cannula. This could result in the need for additional medical intervention to remove the broken piece, pain or loss of m
Which agency issued this recall?
This recall was issued by the FDA Devices on November 13, 2013. Severity: Moderate. Recall number: Z-0127-2014.