PlainRecalls
FDA Devices Moderate Class II Ongoing

FERNO POWER X2 POWERED AMBULANCE COT, Model/Catalog Number: 0015816; FOR USE IN GROUND TRANSPORTATION AMBULANCES.

Reported: October 22, 2025 Initiated: September 11, 2025 #Z-0133-2026

Product Description

FERNO POWER X2 POWERED AMBULANCE COT, Model/Catalog Number: 0015816; FOR USE IN GROUND TRANSPORTATION AMBULANCES.

Reason for Recall

The fastening post hardware could loosen allowing the fastening post to detach from the cot

Details

Recalling Firm
Ferno-Washington Inc
Units Affected
138 units
Distribution
US Nationwide distribution in the states of AL, AR, AZ, CA, CO, FL, IS, IL, KY, LA, MN, MO, NC, ND, NE, NM, NY, OH, OK, PA, TN, TX, UT, VA, WI, WV.
Location
Wilmington, OH

Frequently Asked Questions

What product was recalled?
FERNO POWER X2 POWERED AMBULANCE COT, Model/Catalog Number: 0015816; FOR USE IN GROUND TRANSPORTATION AMBULANCES.. Recalled by Ferno-Washington Inc. Units affected: 138 units.
Why was this product recalled?
The fastening post hardware could loosen allowing the fastening post to detach from the cot
Which agency issued this recall?