Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085

Reported: November 1, 2023 Initiated: September 8, 2023 #Z-0160-2024

Product Description

Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085

Reason for Recall

An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable.

Details

Units Affected
31 units
Distribution
Alabama, California, Illinois, Indiana, Kentucky, Louisiana, Missouri, Montana, North Carolina, Ohio, Oklahoma
Location
Sanford, FL

Frequently Asked Questions

What product was recalled?
Omega Medical Imaging Elevating Monitor Suspension Systems, Part Number 1000-0085. Recalled by Omega Medical Imaging, Inc.. Units affected: 31 units.
Why was this product recalled?
An actuator separated from the pivot mechanism on a lower monitor boom system column. The lower boom became detached at the point shown and was restrained by the installed safety cable.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 1, 2023. Severity: Moderate. Recall number: Z-0160-2024.