FDA Devices recall · Reported November 5, 2025

IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra

There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice …

Recall #
Z-0162-2026
Affected scope
N/A
Initiated
September 30, 2025
Verify with FDA Devices →
View my saved recalls

Impris Imaging Inc recalled IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra — a critical-severity action.

IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra was recalled by Impris Imaging Inc in November 5, 2025. Reason: There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system …. Check the official notice for the remedy. Verify recall #Z-0162-2026 with the FDA Devices before acting.

The recall

Impris Imaging Inc issued this critical-severity FDA Devices recall — There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system ….

Critical
severity level
Class I
classification
November 5, 2025
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0162-2026 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0162-2026) was formally reported on November 5, 2025, with the manufacturer initiating the action on September 30, 2025. It is classified under Critical severity (Class I), with a current status of Ongoing. Impris Imaging Inc is listed as the recalling firm, operating out of Chaska, MN. Federal records list the affected scope as N/A.

The documented reason for this recall is: There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet ven… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices — clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category — 40,409 medical devices recalls on record

-1,00001,0002,0003,0004,0005,000 20052008201120142017202020232026 1,243

Where this recall sits in the database

Severity2366872097High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 100,165 recalls in the database, 23,668 are high severity, 72,097 moderate, and 4,400 low. This recall is classified high severity.

Counts reflect market size and reporting activity, not inherent danger — we do not rank products by risk from raw recall volume.

Severity

Critical

Affected scope

N/A

Related Recalls

6

6 from same agency

Product description

IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra

Reason for recall

There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Ongoing
Recall number Z-0162-2026
Date reported November 5, 2025
Date initiated September 30, 2025
Recalling firm Impris Imaging Inc
Firm location Chaska, MN
Affected scope N/A
Distribution Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0162-2026) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting — recall details can be updated.
  • This is a Class I (high-risk) recall — stop using the product immediately and follow the disposal or return instructions. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
IMRIS Operating Suite, iMRI 3T S, containing Siemens 3 Tesla MRI Magnet, Skyra. Recalled by Impris Imaging Inc. Units affected: N/A.
Why was this product recalled?
There is a potential safety issue associated with the magnet venting paths of the Siemens 3 Tesla MRI system within your IMRIS Operating Suite. Siemens notified IMRIS that an ice blockage may exist within the magnet venting system. In the event of a quench, helium gas may be unable to escape through the designed vent paths, leading to a pressure build-up within the helium containment system which could ultimately rupture the helium containment system, potentially resulting in a helium leak into the scanning room.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 5, 2025. Severity: Critical. Recall number: Z-0162-2026.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Sweden, Switzerland, Japan, Australia, Qatar, and China..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0162-2026) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with GE HealthCare TRADE IN CARESCAPE TELEMETRY SERVER ARK-2250L… →

Data Sources

Source: FDA, CPSC, and NHTSA federal recall databases. This recall: FDA Devices, reported November 5, 2025.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records — no number is typed in by an editor. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of June 2026.