Severity
Moderate
FDA Devices recall · Reported November 9, 2022
Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing i…
Johnson & Johnson Surgical Vision, Inc. recalled THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for … - a moderate-severity action.
THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for … was recalled by Johnson & Johnson Surgical Vision, Inc. in November 9, 2022. Reason: Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path…. Check the official notice for the remedy. Verify recall #Z-0169-2023 with the FDA Devices before acting.
The recall
Johnson & Johnson Surgical Vision, Inc. issued this moderate-severity FDA Devices recall-Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0169-2023 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0169-2023) was formally reported on November 9, 2022, with the manufacturer initiating the action on September 7, 2022. It is classified under Moderate severity (Class II), with a current status of Ongoing. Johnson & Johnson Surgical Vision, Inc. is listed as the recalling firm, operating out of Irvine, CA. Federal records list the affected scope as 91.
The documented reason for this recall is: Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation ca… Distribution data in the federal record shows the product reached: Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
91
Related Recalls
6
3 from same agency
THERMOCOOL SMARTTOUCH SF Bi-Directional Navigation Catheter, REF: D-1348-05-S. Used for catheter-based cardiac electrophysiological mapping (stimulation and recording) and, in conjunction with an RF generator, for cardiac ablation.
Manufacturing error resulted in potential contamination of Bi-Directional navigation catheter irrigation path with cellulose fibers that can cause 1) Failure of initial flushing inspection, 2) Insufficient irrigation causing catheter tip temperature increase, limiting the ability to deliver ablation, or 3) particles flushed through the catheter irrigation may have thrombogenic/toxic effect.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0169-2023 |
| Date reported | November 9, 2022 |
| Date initiated | September 7, 2022 |
| Recalling firm | Johnson & Johnson Surgical Vision, Inc. |
| Firm location | Irvine, CA |
| Affected scope | 91 |
| Distribution | Worldwide - US Nationwide distribution in the states of NC, TN, VA, OH, MO, MI, GA, OR, MO, SD, CO, OH, FL, NJ, TX, IL, WI and the countries of CROATIA, GERMANY, HUNGARY. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 9, 2022. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.