BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel used to test antibiotic susceptibility. The inducible macrolide-linosamide-steptogramin B (iMLSb) resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. --- Product is packaged 25 panels to 1 box.

Reported: December 6, 2017 Initiated: May 11, 2017 #Z-0171-2018

Product Description

BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel used to test antibiotic susceptibility. The inducible macrolide-linosamide-steptogramin B (iMLSb) resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. --- Product is packaged 25 panels to 1 box.

Reason for Recall

BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. As the product should not be used when quality control (QC) fails, it is unlikely that there would be any impact on patient results, however, if a customer does not perform QC testing and uses the panel, there is a potential to incorrectly treat a Staphylococcus infection with clindamycin based on the iMLSb test result. This issue can be detected 100% of the time when used per package insert instructions with iMLSb positive strain S. aureus BAA-977.

Details

Recalling Firm
Becton Dickinson & Co.
Units Affected
US: 7750 units
Distribution
CT, VA, PA, TX, FL, NC, WI, NM
Location
Sparks, MD

Frequently Asked Questions

What product was recalled?
BD Phoenix PMIC-107, Catalog Number 448417, UDI #s (01)30382904484171, (10)7031818(17)180228(30)0025, IVD. PMIC-107 is a Gram positive AST Only panel used to test antibiotic susceptibility. The inducible macrolide-linosamide-steptogramin B (iMLSb) resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. --- Product is packaged 25 panels to 1 box.. Recalled by Becton Dickinson & Co.. Units affected: US: 7750 units.
Why was this product recalled?
BD has recently confirmed quality control failures, related to a manufacturing issue, for the inducible macrolide-linosamide-streptogramin B (iMLSb) test contained on lot 7031818 of Phoenix panels. The iMLSb resistance D-test screen for Staphylococcus species is used to identify phenotypic drug resistance to clindamycin and erythromycin. As the product should not be used when quality control (QC) fails, it is unlikely that there would be any impact on patient results, however, if a customer does not perform QC testing and uses the panel, there is a potential to incorrectly treat a Staphylococcus infection with clindamycin based on the iMLSb test result. This issue can be detected 100% of the time when used per package insert instructions with iMLSb positive strain S. aureus BAA-977.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 6, 2017. Severity: Moderate. Recall number: Z-0171-2018.