Severity
Moderate
FDA Devices recall · Reported November 4, 2015
During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer form the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titani…
Encore Medical, Lp recalled Turon Shoulder Impaction Fixture REF 804-15-102 The Turon and RSP Impaction Fixtures are… - a moderate-severity action.
Turon Shoulder Impaction Fixture REF 804-15-102 The Turon and RSP Impaction Fixtures are… was recalled by Encore Medical, Lp in November 4, 2015. Reason: During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer form the Impaction…. Check the official notice for the remedy. Verify recall #Z-0179-2016 with the FDA Devices before acting.
The recall
Encore Medical, Lp issued this moderate-severity FDA Devices recall-During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer form the Impaction….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0179-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0179-2016) was formally reported on November 4, 2015, with the manufacturer initiating the action on September 14, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Encore Medical, Lp is listed as the recalling firm, operating out of Austin, TX. Federal records list the affected scope as 277 units.
The documented reason for this recall is: During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer form the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating. Distribution data in the federal record shows the product reached: Nationwide Distribution. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
277 units
Related Recalls
6
3 from same agency
Turon Shoulder Impaction Fixture REF 804-15-102 The Turon and RSP Impaction Fixtures are used in the operating room on the back table to assemble the shoulder systems for implant. The Turon Fixture is used to connect the Humeral Head to the Humeral Neck, then subsequently to assemble the Humeral Stem to the Head/Neck construct. The RSP Fixture is used to assemble the UHMWPE Socket Inserts to both the RSP Socket Shells and Monoblock Stems.
During the Turon assembly, the impaction forces caused the polymer, black acetal copolymer form the Impaction Fixture to wear off on the lateral surface of the Humeral Stem titanium plasma spray coating.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0179-2016 |
| Date reported | November 4, 2015 |
| Date initiated | September 14, 2015 |
| Recalling firm | Encore Medical, Lp |
| Firm location | Austin, TX |
| Affected scope | 277 units |
| Distribution | Nationwide Distribution |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 4, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.