Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT TIBIAL "913 PCR" T.1 - EP.10-, PGA00050, Lot Numbers: V11272719, W02300988, W09372781, W09372781, and W10372876 INSERT TIBIAL "913 PCR" T.1 - EP.14-, PGA00054, Lot Number: V0266952 INSERT TIBIAL "913 PCR" T.2 - EP.10-, PGA00080, Lot Numbers: X02396582, X05401377, and W01287345 INSERT TIBIAL "913 PCR" T.3 - EP.10-, PGA00110, Lot Numbers: W09359102, X03396584, and W1166335 INSERT TIBIAL "913 PCR" T.5 -

Reported: November 7, 2012 Initiated: September 5, 2012 #Z-0184-2013 729 units units

Wright Medical Technology Inc issued this FDA Devices recall on November 7, 2012. Classified as Moderate severity (Class II). Approximately 729 units units are affected. The recall was issued because: There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.. This recall notice is sourced from official FDA Devices enforcement records. Below you will find the complete product description, hazard information, remedy instructions, and related recalls from the same manufacturer or product category.

Recall Insight

This FDA Devices action (record #Z-0184-2013) was formally reported on November 7, 2012, with the manufacturer initiating the action on September 5, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Wright Medical Technology Inc is listed as the recalling firm, operating out of Arlington, TN. Federal records indicate 729 units units are affected.

The documented reason for this recall is: There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility. Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) and Internationally to Canada, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Lithuania, Norway, Poland, Portug…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

To put this record in context, PlainRecalls indexes 83,949 recalls across the FDA, CPSC, NHTSA and USDA FSIS going back to 1995. Within the same product category, the database holds 6 closely related recalls, of which 6 were also issued by FDA Devices. That clustering is a signal — repeated actions in a narrow category often indicate a systemic quality-control issue, a supplier-wide contamination, or a design defect that has propagated across product lines. This recall is roughly 14 years old; older recalls can remain relevant because many units enter resale, rental, and secondary-market channels where the original warning never reaches the end user. Always cross-check the recall number against the official agency page before relying on any summary.

Recall Distribution by Severity Class

Severity1Class I (Critical)Class II (Moderate)Class III (Low)
Recall Distribution by Severity Class

Severity

Moderate

Units Affected

729 units

Related Recalls

6

6 from same agency

Product Description

Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT TIBIAL "913 PCR" T.1 - EP.10-, PGA00050, Lot Numbers: V11272719, W02300988, W09372781, W09372781, and W10372876 INSERT TIBIAL "913 PCR" T.1 - EP.14-, PGA00054, Lot Number: V0266952 INSERT TIBIAL "913 PCR" T.2 - EP.10-, PGA00080, Lot Numbers: X02396582, X05401377, and W01287345 INSERT TIBIAL "913 PCR" T.3 - EP.10-, PGA00110, Lot Numbers: W09359102, X03396584, and W1166335 INSERT TIBIAL "913 PCR" T.5 - EP.14-, PGA00174, Lot Number: W0675535 PLATEAU TIBIAL "913" T.4 MOINS , PGA00216, Lot Number: X10391238 PLATEAU TIBIAL "913" T.5 MOINS , PGA00218, Lot Numbers: V0379902 and V0381596 PLATEAU TIBIAL "913" T.1 , PGA00230, Lot Numbers: W05327837 and W0779903 PLATEAU TIBIAL "913" T.2 PLUS , PGA00252, Lot Number: X06391241 TIGE INTRAMEDULLAIRE "913" D=16 L=75, PGA00716, Lot Number: X02382967 TIGE INTRAMEDULLAIRE "913" D=12 L=125, PGA00752, Lot Number: W06322599 TIGE INTRAMEDULLAIRE "913" D=14 L=125, PGA00754, Lot Number: V12272733 TIGE INTRAMEDULLAIRE "913" D=16 L=125, PGA00756, Lot Number: V12272734 TIGE INTRAMEDULLAIRE "913" D=18 L=125, PGA00758, Lot Number: W04336344 TIGE DE REPRISE FEMO/TIB. "913" D=14 L=75, PGA00814, Lot Number: V0767876 TIGE DE REPRISE FEMO/TIB. "913" D=16 L=75, PGA00816, Lot Number: V0767877 CALE TIBIALE "913" T.1 -EP.8- , PGA00950, Lot Number: W0380839 CALE TIBIALE "913" T.2 -EP.8- , PGA00952, Lot Number: V0773498 and V0973301 INSERT TIBIAL "913 RS" T.1 - EP.10-, PGA14010, Lot Number: V0782533 INSERT TIBIAL "913 RS" T.1 - EP.12-, PGA14012, Lot Number: V0766107 and V0782536 INSERT TIBIAL "913 RS" T.1 - EP.16-, PGA14016, V0782531 INSERT TIBIAL "913 RS" T.5 - EP.12-, PGA14044, Lot Number V0782542 INSERT TIBIAL "913 RS" T.5 - EP.14-, PGA14046, Lot Number: V0782540 INSERT TIBIAL "913 MP" T.1 GAUCHE -EP.17-, PGA14508, Lot Number: V05210017 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 38, PGA20016, Lot Numbers: X03397144, X04401212, X05401380, X05418562, and X05419556 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 41, PGA20018, Lot Numbers: X03397145, X04400214, X04401213, X05401381, and X05418563 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 35, PGA20024, Lot Number: X07429260 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 38, PGA20026, Lot Numbers: X03397148, X04400217, X04401217, X05401384, and X05418557 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 41, PGA20028, Lot Numbers: X03397149, X04400218, X04401218, X05401385, and X05418559 TROCHLEA "FPV EVOS" RIGHT SIZE 3, PGA20116, Lot Number: X05427577 TROCHLEA "FPV EVOS" RIGHT SIZE 4, PGA20118, Lot Numbers: X03397138, X05427571, X05427578, and X07428738 TROCHLEA "FPV EVOS" LEFT SIZE 3, PGA20126, Lot Numbers: X05427573 and X05427582 TROCHLEA "FPV EVOS" LEFT SIZE 4, PGA20128, Lot Numbers: X03397141, X05427567, X05427574, X05427583, X0542758310, and X07427587 PLAT. TIBIAL TITANE IB II-59 S/CIMENT EQUIPE, PPB21002, Lot Number: W11373017 PLAT. TIBIAL TITANE IB II-69 S/CIMENT EQUIPE, PPB21006, Lot Number: W01278824 PLAT. TIBIAL TITANE IB II-74 S/CIMENT EQUIPE, PPB21008, Lot Numbers: V0281306 and V1128727 2 VIS POUR PLATEAU TIBIAL S/C IB II:D=5.5*20 (TITANE), PPB22000, Lot Number: X04418520 PLAT. TIBIAL TITANE IB II-54 A/CIMENT EQUIPE, PPB30030, Lot Number: W09359212 PLAT. TIBIAL TITANE IB II-59 A/CIMENT EQUIPE, PPB30034, Lot Number: W11373000 PLATEAU TIBIAL PE IB II-64 8.5, PPB30140, Lot Numbers: V12278842, W01301110, W03308823, W05327953, W07350813, and W09359226 PLATEAU TIBIAL PE IB II-64 10, PPB30142, Lot Numbers: V11272779 and W01288273 PLATEAU TIBIAL PE IB II-64 12.5, PPB30144, Lot Numbers: V09243783, W01278844, and W09359228 PLATEAU TIBIAL PE IB II-64 17.5, PPB30148, Lot Numbers: W06335185 and W06344258 PLATEAU TIBIAL PE IB II-64 20, PPB30150, Lot Number: V05194572 PLATEAU TIBIAL PE IB II-69 8.5, PPB30160, Lot Numbers: V11272780 and W01278845 PLATEAU TIBIAL PE IB II-69 10, PPB30162, Lot Numbers: V09243785, V09244054, and W01278846 PLATEAU TIBI

Reason for Recall

There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.

Details

Units Affected
729 units
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to Canada, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Lithuania, Norway, Poland, Portugal, Serbia, Slovakia, Slovenia, Spain, The Netherlands, Argentina, Brazil, Chile, China, Japan, Taiwan, Turkey, United Arab Emirates, Australia, South Africa, Russia, and United Kingdom.
Location
Arlington, TN

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0184-2013
Date reported November 7, 2012
Date initiated September 5, 2012
Recalling firm Wright Medical Technology Inc
Units affected 729 units
Distribution Worldwide Distribution - USA (nationwide) and Internationally to Canada, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Lithuania, Norway, Poland, Portugal, Serbia, Slovakia…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

729 units units affected — limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units)
Large-scale (100K – 1M units)
Massive (≥1M units)

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

Frequently Asked Questions

What product was recalled?
Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT TIBIAL "913 PCR" T.1 - EP.10-, PGA00050, Lot Numbers: V11272719, W02300988, W09372781, W09372781, and W10372876 INSERT TIBIAL "913 PCR" T.1 - EP.14-, PGA00054, Lot Number: V0266952 INSERT TIBIAL "913 PCR" T.2 - EP.10-, PGA00080, Lot Numbers: X02396582, X05401377, and W01287345 INSERT TIBIAL "913 PCR" T.3 - EP.10-, PGA00110, Lot Numbers: W09359102, X03396584, and W1166335 INSERT TIBIAL "913 PCR" T.5 - EP.14-, PGA00174, Lot Number: W0675535 PLATEAU TIBIAL "913" T.4 MOINS , PGA00216, Lot Number: X10391238 PLATEAU TIBIAL "913" T.5 MOINS , PGA00218, Lot Numbers: V0379902 and V0381596 PLATEAU TIBIAL "913" T.1 , PGA00230, Lot Numbers: W05327837 and W0779903 PLATEAU TIBIAL "913" T.2 PLUS , PGA00252, Lot Number: X06391241 TIGE INTRAMEDULLAIRE "913" D=16 L=75, PGA00716, Lot Number: X02382967 TIGE INTRAMEDULLAIRE "913" D=12 L=125, PGA00752, Lot Number: W06322599 TIGE INTRAMEDULLAIRE "913" D=14 L=125, PGA00754, Lot Number: V12272733 TIGE INTRAMEDULLAIRE "913" D=16 L=125, PGA00756, Lot Number: V12272734 TIGE INTRAMEDULLAIRE "913" D=18 L=125, PGA00758, Lot Number: W04336344 TIGE DE REPRISE FEMO/TIB. "913" D=14 L=75, PGA00814, Lot Number: V0767876 TIGE DE REPRISE FEMO/TIB. "913" D=16 L=75, PGA00816, Lot Number: V0767877 CALE TIBIALE "913" T.1 -EP.8- , PGA00950, Lot Number: W0380839 CALE TIBIALE "913" T.2 -EP.8- , PGA00952, Lot Number: V0773498 and V0973301 INSERT TIBIAL "913 RS" T.1 - EP.10-, PGA14010, Lot Number: V0782533 INSERT TIBIAL "913 RS" T.1 - EP.12-, PGA14012, Lot Number: V0766107 and V0782536 INSERT TIBIAL "913 RS" T.1 - EP.16-, PGA14016, V0782531 INSERT TIBIAL "913 RS" T.5 - EP.12-, PGA14044, Lot Number V0782542 INSERT TIBIAL "913 RS" T.5 - EP.14-, PGA14046, Lot Number: V0782540 INSERT TIBIAL "913 MP" T.1 GAUCHE -EP.17-, PGA14508, Lot Number: V05210017 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 38, PGA20016, Lot Numbers: X03397144, X04401212, X05401380, X05418562, and X05419556 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 41, PGA20018, Lot Numbers: X03397145, X04400214, X04401213, X05401381, and X05418563 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 35, PGA20024, Lot Number: X07429260 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 38, PGA20026, Lot Numbers: X03397148, X04400217, X04401217, X05401384, and X05418557 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 41, PGA20028, Lot Numbers: X03397149, X04400218, X04401218, X05401385, and X05418559 TROCHLEA "FPV EVOS" RIGHT SIZE 3, PGA20116, Lot Number: X05427577 TROCHLEA "FPV EVOS" RIGHT SIZE 4, PGA20118, Lot Numbers: X03397138, X05427571, X05427578, and X07428738 TROCHLEA "FPV EVOS" LEFT SIZE 3, PGA20126, Lot Numbers: X05427573 and X05427582 TROCHLEA "FPV EVOS" LEFT SIZE 4, PGA20128, Lot Numbers: X03397141, X05427567, X05427574, X05427583, X0542758310, and X07427587 PLAT. TIBIAL TITANE IB II-59 S/CIMENT EQUIPE, PPB21002, Lot Number: W11373017 PLAT. TIBIAL TITANE IB II-69 S/CIMENT EQUIPE, PPB21006, Lot Number: W01278824 PLAT. TIBIAL TITANE IB II-74 S/CIMENT EQUIPE, PPB21008, Lot Numbers: V0281306 and V1128727 2 VIS POUR PLATEAU TIBIAL S/C IB II:D=5.5*20 (TITANE), PPB22000, Lot Number: X04418520 PLAT. TIBIAL TITANE IB II-54 A/CIMENT EQUIPE, PPB30030, Lot Number: W09359212 PLAT. TIBIAL TITANE IB II-59 A/CIMENT EQUIPE, PPB30034, Lot Number: W11373000 PLATEAU TIBIAL PE IB II-64 8.5, PPB30140, Lot Numbers: V12278842, W01301110, W03308823, W05327953, W07350813, and W09359226 PLATEAU TIBIAL PE IB II-64 10, PPB30142, Lot Numbers: V11272779 and W01288273 PLATEAU TIBIAL PE IB II-64 12.5, PPB30144, Lot Numbers: V09243783, W01278844, and W09359228 PLATEAU TIBIAL PE IB II-64 17.5, PPB30148, Lot Numbers: W06335185 and W06344258 PLATEAU TIBIAL PE IB II-64 20, PPB30150, Lot Number: V05194572 PLATEAU TIBIAL PE IB II-69 8.5, PPB30160, Lot Numbers: V11272780 and W01278845 PLATEAU TIBIAL PE IB II-69 10, PPB30162, Lot Numbers: V09243785, V09244054, and W01278846 PLATEAU TIBI. Recalled by Wright Medical Technology Inc. Units affected: 729 units.
Why was this product recalled?
There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 7, 2012. Severity: Moderate. Recall number: Z-0184-2013.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - USA (nationwide) and Internationally to Canada, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Lithuania, Norway, Poland, Portugal, Serbia, Slovakia, Slovenia, Spain, The Netherlands, Argentina, Brazil, Chile, China, Japan, Taiwan, Turkey, United Arab Emirates, Australia, South Africa, Russia, and United Kingdom..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0184-2013) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Recall Context

Product recalls are issued when a manufacturer, distributor, or federal agency determines that a product poses a safety risk to consumers. This recall is classified as moderate severity, indicating the product may cause temporary or medically reversible health consequences. Across PlainRecalls, we track 83,000+ recalls from FDA, CPSC, and NHTSA to help consumers stay informed and act quickly when safety issues arise.

Nearby Recalls in This Category

Other recalls in the same product category — useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Endo-Model Replacement Plateau; Item Number: 15-0027/11; →

Data Sources

Data as of 2025. Source: FDA, CPSC, NHTSA, USDA FSIS federal recall databases.

  • Source: FDA — Food and Drug Administration, openFDA Enforcement API (food, drug, and medical device recalls)
  • Source: CPSC — Consumer Product Safety Commission Recalls API (consumer product recalls and hazards)
  • Source: NHTSA — National Highway Traffic Safety Administration Recalls API (vehicle safety recalls)
  • Source: USDA FSIS — Food Safety and Inspection Service (meat, poultry, and egg product recalls)

Recall information is sourced from official federal agency databases. Always verify recall details with the issuing agency for the most current status. This information is for research and awareness purposes only.

All federal data sources used on this page

Source: Federal recall agencies (FDA, CPSC, NHTSA, USDA FSIS) Aggregated multi-agency recall feeds · 2024 Recall data normalized across federal agency feeds; severity classifications follow each agency's own taxonomy (FDA Class I/II/III; CPSC, NHTSA, USDA FSIS).