PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported November 7, 2012

Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT TIBIAL "913 PCR" T.1 - EP.10-, PGA00050, Lot Numbers: V11272719, W02300988, W09372781, W09372781, and W10372876 INSERT TIBIAL "913 PCR" T.1 - EP.14-, PGA00054, Lot Number: V0266952 INSERT TIBIAL "913 PCR" T.2 - EP.10-, PGA00080, Lot Numbers: X02396582, X05401377, and W01287345 INSERT TIBIAL "913 PCR" T.3 - EP.10-, PGA00110, Lot Numbers: W09359102, X03396584, and W1166335 INSERT TIBIAL "913 PCR" T.5 -

There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.

Recall #
Z-0184-2013
Affected scope
729 units
Initiated
September 5, 2012
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Wright Medical Technology Inc recalled Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT … - a moderate-severity action.

Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT … was recalled by Wright Medical Technology Inc in November 7, 2012. Reason: There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.. Check the official notice for the remedy. Verify recall #Z-0184-2013 with the FDA Devices before acting.

The recall

Wright Medical Technology Inc issued this moderate-severity FDA Devices recall-There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility..

Moderate
severity level
729 units
affected scope
Class II
classification
November 7, 2012
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0184-2013 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0184-2013) was formally reported on November 7, 2012, with the manufacturer initiating the action on September 5, 2012. It is classified under Moderate severity (Class II), with a current status of Terminated. Wright Medical Technology Inc is listed as the recalling firm, operating out of Arlington, TN. Federal records list the affected scope as 729 units.

The documented reason for this recall is: There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility. Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide) and Internationally to Canada, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Lithuania, Norway, Poland, Portug…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

729 units

Related Recalls

6

3 from same agency

Product description

Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT TIBIAL "913 PCR" T.1 - EP.10-, PGA00050, Lot Numbers: V11272719, W02300988, W09372781, W09372781, and W10372876 INSERT TIBIAL "913 PCR" T.1 - EP.14-, PGA00054, Lot Number: V0266952 INSERT TIBIAL "913 PCR" T.2 - EP.10-, PGA00080, Lot Numbers: X02396582, X05401377, and W01287345 INSERT TIBIAL "913 PCR" T.3 - EP.10-, PGA00110, Lot Numbers: W09359102, X03396584, and W1166335 INSERT TIBIAL "913 PCR" T.5 - EP.14-, PGA00174, Lot Number: W0675535 PLATEAU TIBIAL "913" T.4 MOINS , PGA00216, Lot Number: X10391238 PLATEAU TIBIAL "913" T.5 MOINS , PGA00218, Lot Numbers: V0379902 and V0381596 PLATEAU TIBIAL "913" T.1 , PGA00230, Lot Numbers: W05327837 and W0779903 PLATEAU TIBIAL "913" T.2 PLUS , PGA00252, Lot Number: X06391241 TIGE INTRAMEDULLAIRE "913" D=16 L=75, PGA00716, Lot Number: X02382967 TIGE INTRAMEDULLAIRE "913" D=12 L=125, PGA00752, Lot Number: W06322599 TIGE INTRAMEDULLAIRE "913" D=14 L=125, PGA00754, Lot Number: V12272733 TIGE INTRAMEDULLAIRE "913" D=16 L=125, PGA00756, Lot Number: V12272734 TIGE INTRAMEDULLAIRE "913" D=18 L=125, PGA00758, Lot Number: W04336344 TIGE DE REPRISE FEMO/TIB. "913" D=14 L=75, PGA00814, Lot Number: V0767876 TIGE DE REPRISE FEMO/TIB. "913" D=16 L=75, PGA00816, Lot Number: V0767877 CALE TIBIALE "913" T.1 -EP.8- , PGA00950, Lot Number: W0380839 CALE TIBIALE "913" T.2 -EP.8- , PGA00952, Lot Number: V0773498 and V0973301 INSERT TIBIAL "913 RS" T.1 - EP.10-, PGA14010, Lot Number: V0782533 INSERT TIBIAL "913 RS" T.1 - EP.12-, PGA14012, Lot Number: V0766107 and V0782536 INSERT TIBIAL "913 RS" T.1 - EP.16-, PGA14016, V0782531 INSERT TIBIAL "913 RS" T.5 - EP.12-, PGA14044, Lot Number V0782542 INSERT TIBIAL "913 RS" T.5 - EP.14-, PGA14046, Lot Number: V0782540 INSERT TIBIAL "913 MP" T.1 GAUCHE -EP.17-, PGA14508, Lot Number: V05210017 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 38, PGA20016, Lot Numbers: X03397144, X04401212, X05401380, X05418562, and X05419556 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 41, PGA20018, Lot Numbers: X03397145, X04400214, X04401213, X05401381, and X05418563 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 35, PGA20024, Lot Number: X07429260 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 38, PGA20026, Lot Numbers: X03397148, X04400217, X04401217, X05401384, and X05418557 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 41, PGA20028, Lot Numbers: X03397149, X04400218, X04401218, X05401385, and X05418559 TROCHLEA "FPV EVOS" RIGHT SIZE 3, PGA20116, Lot Number: X05427577 TROCHLEA "FPV EVOS" RIGHT SIZE 4, PGA20118, Lot Numbers: X03397138, X05427571, X05427578, and X07428738 TROCHLEA "FPV EVOS" LEFT SIZE 3, PGA20126, Lot Numbers: X05427573 and X05427582 TROCHLEA "FPV EVOS" LEFT SIZE 4, PGA20128, Lot Numbers: X03397141, X05427567, X05427574, X05427583, X0542758310, and X07427587 PLAT. TIBIAL TITANE IB II-59 S/CIMENT EQUIPE, PPB21002, Lot Number: W11373017 PLAT. TIBIAL TITANE IB II-69 S/CIMENT EQUIPE, PPB21006, Lot Number: W01278824 PLAT. TIBIAL TITANE IB II-74 S/CIMENT EQUIPE, PPB21008, Lot Numbers: V0281306 and V1128727 2 VIS POUR PLATEAU TIBIAL S/C IB II:D=5.5*20 (TITANE), PPB22000, Lot Number: X04418520 PLAT. TIBIAL TITANE IB II-54 A/CIMENT EQUIPE, PPB30030, Lot Number: W09359212 PLAT. TIBIAL TITANE IB II-59 A/CIMENT EQUIPE, PPB30034, Lot Number: W11373000 PLATEAU TIBIAL PE IB II-64 8.5, PPB30140, Lot Numbers: V12278842, W01301110, W03308823, W05327953, W07350813, and W09359226 PLATEAU TIBIAL PE IB II-64 10, PPB30142, Lot Numbers: V11272779 and W01288273 PLATEAU TIBIAL PE IB II-64 12.5, PPB30144, Lot Numbers: V09243783, W01278844, and W09359228 PLATEAU TIBIAL PE IB II-64 17.5, PPB30148, Lot Numbers: W06335185 and W06344258 PLATEAU TIBIAL PE IB II-64 20, PPB30150, Lot Number: V05194572 PLATEAU TIBIAL PE IB II-69 8.5, PPB30160, Lot Numbers: V11272780 and W01278845 PLATEAU TIBIAL PE IB II-69 10, PPB30162, Lot Numbers: V09243785, V09244054, and W01278846 PLATEAU TIBI

Reason for recall

There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0184-2013
Date reported November 7, 2012
Date initiated September 5, 2012
Recalling firm Wright Medical Technology Inc
Firm location Arlington, TN
Affected scope 729 units
Distribution Worldwide Distribution - USA (nationwide) and Internationally to Canada, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Lithuania, Norway, Poland, Portugal, Serbia, Slovakia…

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

729 units units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0184-2013) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Knee Components: ROTULE "913" D=32 , Part Number PGA00030, Lot Number X02383380 INSERT TIBIAL "913 PCR" T.1 - EP.10-, PGA00050, Lot Numbers: V11272719, W02300988, W09372781, W09372781, and W10372876 INSERT TIBIAL "913 PCR" T.1 - EP.14-, PGA00054, Lot Number: V0266952 INSERT TIBIAL "913 PCR" T.2 - EP.10-, PGA00080, Lot Numbers: X02396582, X05401377, and W01287345 INSERT TIBIAL "913 PCR" T.3 - EP.10-, PGA00110, Lot Numbers: W09359102, X03396584, and W1166335 INSERT TIBIAL "913 PCR" T.5 - EP.14-, PGA00174, Lot Number: W0675535 PLATEAU TIBIAL "913" T.4 MOINS , PGA00216, Lot Number: X10391238 PLATEAU TIBIAL "913" T.5 MOINS , PGA00218, Lot Numbers: V0379902 and V0381596 PLATEAU TIBIAL "913" T.1 , PGA00230, Lot Numbers: W05327837 and W0779903 PLATEAU TIBIAL "913" T.2 PLUS , PGA00252, Lot Number: X06391241 TIGE INTRAMEDULLAIRE "913" D=16 L=75, PGA00716, Lot Number: X02382967 TIGE INTRAMEDULLAIRE "913" D=12 L=125, PGA00752, Lot Number: W06322599 TIGE INTRAMEDULLAIRE "913" D=14 L=125, PGA00754, Lot Number: V12272733 TIGE INTRAMEDULLAIRE "913" D=16 L=125, PGA00756, Lot Number: V12272734 TIGE INTRAMEDULLAIRE "913" D=18 L=125, PGA00758, Lot Number: W04336344 TIGE DE REPRISE FEMO/TIB. "913" D=14 L=75, PGA00814, Lot Number: V0767876 TIGE DE REPRISE FEMO/TIB. "913" D=16 L=75, PGA00816, Lot Number: V0767877 CALE TIBIALE "913" T.1 -EP.8- , PGA00950, Lot Number: W0380839 CALE TIBIALE "913" T.2 -EP.8- , PGA00952, Lot Number: V0773498 and V0973301 INSERT TIBIAL "913 RS" T.1 - EP.10-, PGA14010, Lot Number: V0782533 INSERT TIBIAL "913 RS" T.1 - EP.12-, PGA14012, Lot Number: V0766107 and V0782536 INSERT TIBIAL "913 RS" T.1 - EP.16-, PGA14016, V0782531 INSERT TIBIAL "913 RS" T.5 - EP.12-, PGA14044, Lot Number V0782542 INSERT TIBIAL "913 RS" T.5 - EP.14-, PGA14046, Lot Number: V0782540 INSERT TIBIAL "913 MP" T.1 GAUCHE -EP.17-, PGA14508, Lot Number: V05210017 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 38, PGA20016, Lot Numbers: X03397144, X04401212, X05401380, X05418562, and X05419556 PATELLA TRI-PEG "FPV EVOS" RIGHT SIZE 41, PGA20018, Lot Numbers: X03397145, X04400214, X04401213, X05401381, and X05418563 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 35, PGA20024, Lot Number: X07429260 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 38, PGA20026, Lot Numbers: X03397148, X04400217, X04401217, X05401384, and X05418557 PATELLA TRI-PEG "FPV EVOS" LEFT SIZE 41, PGA20028, Lot Numbers: X03397149, X04400218, X04401218, X05401385, and X05418559 TROCHLEA "FPV EVOS" RIGHT SIZE 3, PGA20116, Lot Number: X05427577 TROCHLEA "FPV EVOS" RIGHT SIZE 4, PGA20118, Lot Numbers: X03397138, X05427571, X05427578, and X07428738 TROCHLEA "FPV EVOS" LEFT SIZE 3, PGA20126, Lot Numbers: X05427573 and X05427582 TROCHLEA "FPV EVOS" LEFT SIZE 4, PGA20128, Lot Numbers: X03397141, X05427567, X05427574, X05427583, X0542758310, and X07427587 PLAT. TIBIAL TITANE IB II-59 S/CIMENT EQUIPE, PPB21002, Lot Number: W11373017 PLAT. TIBIAL TITANE IB II-69 S/CIMENT EQUIPE, PPB21006, Lot Number: W01278824 PLAT. TIBIAL TITANE IB II-74 S/CIMENT EQUIPE, PPB21008, Lot Numbers: V0281306 and V1128727 2 VIS POUR PLATEAU TIBIAL S/C IB II:D=5.5*20 (TITANE), PPB22000, Lot Number: X04418520 PLAT. TIBIAL TITANE IB II-54 A/CIMENT EQUIPE, PPB30030, Lot Number: W09359212 PLAT. TIBIAL TITANE IB II-59 A/CIMENT EQUIPE, PPB30034, Lot Number: W11373000 PLATEAU TIBIAL PE IB II-64 8.5, PPB30140, Lot Numbers: V12278842, W01301110, W03308823, W05327953, W07350813, and W09359226 PLATEAU TIBIAL PE IB II-64 10, PPB30142, Lot Numbers: V11272779 and W01288273 PLATEAU TIBIAL PE IB II-64 12.5, PPB30144, Lot Numbers: V09243783, W01278844, and W09359228 PLATEAU TIBIAL PE IB II-64 17.5, PPB30148, Lot Numbers: W06335185 and W06344258 PLATEAU TIBIAL PE IB II-64 20, PPB30150, Lot Number: V05194572 PLATEAU TIBIAL PE IB II-69 8.5, PPB30160, Lot Numbers: V11272780 and W01278845 PLATEAU TIBIAL PE IB II-69 10, PPB30162, Lot Numbers: V09243785, V09244054, and W01278846 PLATEAU TIBI. Recalled by Wright Medical Technology Inc. Units affected: 729 units.
Why was this product recalled?
There may be inaccuracies on the labeled shelf life on all products packaged at the Toulon, France facility.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 7, 2012. Severity: Moderate. Recall number: Z-0184-2013.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - USA (nationwide) and Internationally to Canada, Austria, Belgium, Bulgaria, Cyprus, Czech Republic, Denmark, Finland, France, Germany, Greece, Italy, Latvia, Lithuania, Norway, Poland, Portugal, Serbia, Slovakia, Slovenia, Spain, The Netherlands, Argentina, Brazil, Chile, China, Japan, Taiwan, Turkey, United Arab Emirates, Australia, South Africa, Russia, and United Kingdom..
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0184-2013) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 7, 2012. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.