pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per blister, 12 blisters per box), STERILE R, Rx ONLY The DORO(R) Skull Pins are inserted into the receptacles of the skull clamp and fixate the head during surgery.

The firm received reports of two breakages of the pin tips which occurred during two identical surgery procedures. The breakage occurred at the very distal end of the conical par…

Recall #
Z-0193-2018
Affected scope
1142 units
Initiated
November 2, 2017
Compiled from official public sources by the editorial team.
Verify with FDA Devices →
View my saved recalls

Pro-Med Instruments Gmbh recalled pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per b… - a moderate-severity action.

pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per b… was recalled by Pro-Med Instruments Gmbh in December 13, 2017. Reason: The firm received reports of two breakages of the pin tips which occurred during two identical surgery proced…. Check the official notice for the remedy. Verify recall #Z-0193-2018 with the FDA Devices before acting.

The recall

Pro-Med Instruments Gmbh issued this moderate-severity FDA Devices recall-The firm received reports of two breakages of the pin tips which occurred during two identical surgery proced….

Moderate
severity level
1K units
affected scope
Class II
classification
December 13, 2017
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0193-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0193-2018) was formally reported on December 13, 2017, with the manufacturer initiating the action on November 2, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Pro-Med Instruments Gmbh is listed as the recalling firm, operating out of Freiburg Im Breisgau, N/A. Federal records list the affected scope as 1142 units.

The documented reason for this recall is: The firm received reports of two breakages of the pin tips which occurred during two identical surgery procedures. The breakage occurred at the very distal end of the conical part of the skull pin tip. Distribution data in the federal record shows the product reached: Worldwide Distribution - US (nationwide) Internationally to Germany, Saudi Arabia, France, United Kingdom, Australia, Austria, Ireland, South Africa, Hungary, India, Spain, Italy, Japan, and Canada. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,322 medical devices recalls on record

-1,00001,0002,0003,0004,000 20052008201120142017202020232026 1,248
Severity2487872426High severity (most serious)Moderate severityLow severity
Where this recall sits in the database

Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Severity

Moderate

Affected scope

1142 units

Related Recalls

6

3 from same agency

Product description

pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per blister, 12 blisters per box), STERILE R, Rx ONLY The DORO(R) Skull Pins are inserted into the receptacles of the skull clamp and fixate the head during surgery.

Reason for recall

The firm received reports of two breakages of the pin tips which occurred during two identical surgery procedures. The breakage occurred at the very distal end of the conical part of the skull pin tip.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0193-2018
Date reported December 13, 2017
Date initiated November 2, 2017
Recalling firm Pro-Med Instruments Gmbh
Firm location Freiburg Im Breisgau, N/A
Affected scope 1142 units
Distribution Worldwide Distribution - US (nationwide) Internationally to Germany, Saudi Arabia, France, United Kingdom, Australia, Austria, Ireland, South Africa, Hungary, India, Spain, Italy, Japan, and Canada

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

1142 units units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0193-2018) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
pmi, pro med instruments, DORO(R) Sterile Disposable Skull Pins, 3 pieces (3 pieces per blister, 12 blisters per box), STERILE R, Rx ONLY The DORO(R) Skull Pins are inserted into the receptacles of the skull clamp and fixate the head during surgery.. Recalled by Pro-Med Instruments Gmbh. Units affected: 1142 units.
Why was this product recalled?
The firm received reports of two breakages of the pin tips which occurred during two identical surgery procedures. The breakage occurred at the very distal end of the conical part of the skull pin tip.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 13, 2017. Severity: Moderate. Recall number: Z-0193-2018.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - US (nationwide) Internationally to Germany, Saudi Arabia, France, United Kingdom, Australia, Austria, Ireland, South Africa, Hungary, India, Spain, Italy, Japan, and Canada.
How do I check if my product is affected by a recall?
Check the product description and recall number (Z-0193-2018) against your product. Visit the official FDA Devices website for the most current information. You can also use our Recall Checker tool to search by product name or brand.
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nearby Recalls in This Category

Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.

Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →

Browse all Medical Devices recalls →

Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 13, 2017. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.