FDA Devices recall · Reported November 11, 2015
1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and wit…
Philips Medical Systems, Inc. recalled Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A - a moderate-severity action.
Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A was recalled by Philips Medical Systems, Inc. in November 11, 2015. Reason: 1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade…. Check the official notice for the remedy. Verify recall #Z-0204-2016 with the FDA Devices before acting.
The recall
Philips Medical Systems, Inc. issued this moderate-severity FDA Devices recall-1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0204-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0204-2016) was formally reported on November 11, 2015, with the manufacturer initiating the action on November 19, 2014. It is classified under Moderate severity (Class II), with a current status of Terminated. Philips Medical Systems, Inc. is listed as the recalling firm, operating out of Andover, MA. Federal records list the affected scope as 1553 units.
The documented reason for this recall is: 1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery in… Distribution data in the federal record shows the product reached: Worldwide Distribution - USA (nationwide), Australia, Austria, Bahrain, Belgium, Brunei Darussalam, Canada, China, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Republic of Korea,…. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
1553 units
Related Recalls
6
3 from same agency
Philips HeartStart MRx Monitor/Defibrillators Models: M3535A and M3536A
1. Device will perform the weekly automated tests hourly, which could cause the therapy capacitors to degrade sooner than intended and 2. While connected to AC or DC power and with no battery installed or the battery installed has a charge level of less than 10%, the Ready for Use (RFU) indicator will not provide the expected low battery indication
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0204-2016 |
| Date reported | November 11, 2015 |
| Date initiated | November 19, 2014 |
| Recalling firm | Philips Medical Systems, Inc. |
| Firm location | Andover, MA |
| Affected scope | 1553 units |
| Distribution | Worldwide Distribution - USA (nationwide), Australia, Austria, Bahrain, Belgium, Brunei Darussalam, Canada, China, Finland, France, Germany, Hong Kong, India, Indonesia, Ireland, Israel, Italy, Japan, Republic of Korea, Libyan Arab Jamahir… |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 11, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.