Mitek Screw and Washer Depth Gauge, Product Code 219024
Reported: October 31, 2018 Initiated: August 30, 2018 #Z-0208-2019
Product Description
Mitek Screw and Washer Depth Gauge, Product Code 219024
Reason for Recall
The device has an offset in the design that results in a reading approximately 3.2mm shorter than the actual length of the bone hole.
Details
- Recalling Firm
- DePuy Mitek, Inc., a Johnson & Johnson Co.
- Units Affected
- 401
- Distribution
- US nationwide in the states of AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, KS, KY, LA, MA, MI, MO, MT, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, TN, TX, UT, VA, VT, WA, WI. The products were distributed to the following foreign countries: Australia, Canada, Chile, China, Chechnya, Denmark, Italy, Japan, Jordan, Netherlands, Poland, Slovenia, Sweden, Turkey, UK.
- Location
- Raynham, MA
Frequently Asked Questions
What product was recalled? ▼
Mitek Screw and Washer Depth Gauge, Product Code 219024. Recalled by DePuy Mitek, Inc., a Johnson & Johnson Co.. Units affected: 401.
Why was this product recalled? ▼
The device has an offset in the design that results in a reading approximately 3.2mm shorter than the actual length of the bone hole.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on October 31, 2018. Severity: Moderate. Recall number: Z-0208-2019.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11