2 Gang 4-Way NanoClave Stopcock Manifold w/Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709053172(17)241201(30)50(10)4493459;(01)10887709053172(17)241201(30)50(10)4536673. The ICU Medical Clave Stopcock is a device used to administer or withdraw fluids that flow from a container to a patient's vascular system through a needle or catheter inserted into a vein or artery. This device is a four-way stopcock with an integral CLAVE side-port for closed, needleless ac
Product Description
2 Gang 4-Way NanoClave Stopcock Manifold w/Rotating Luer; 1 unit per pouch, 50 multivac pouches per case. UDI:(01)10887709053172(17)241201(30)50(10)4493459;(01)10887709053172(17)241201(30)50(10)4536673. The ICU Medical Clave Stopcock is a device used to administer or withdraw fluids that flow from a container to a patient's vascular system through a needle or catheter inserted into a vein or artery. This device is a four-way stopcock with an integral CLAVE side-port for closed, needleless access to the patient tubing or other fluid container.
Reason for Recall
Identification of a potential manufacturing defect on the internal surface of the NanoClave within specific lots of NanoClave sets, which may inhibit a proper seal with the NanoClave spike.
Details
- Recalling Firm
- ICU Medical, Inc.
- Units Affected
- Total of all products (Listed #1 thru 101) = 304735 units
- Distribution
- Worldwide Distribution: US (nationwide): AL, AZ, CA, CO, DE, FL,GA,ID, IL,IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NH,NJ,NM,NV, NY, OH, OK, OR, PA,RI, TN, TX, UT, VA, WA, WI, and WV; and OUS countries of: Belgium, Brazil, Canada, France, Germany, Ireland, Japan, Martinique, Mexico, Poland, and Saudi Arabia.
- Location
- San Clemente, CA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11