Severity
Moderate
FDA Devices recall · Reported November 15, 2023
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may…
Magellan Diagnostics, Inc. recalled LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the… - a moderate-severity action.
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the… was recalled by Magellan Diagnostics, Inc. in November 15, 2023. Reason: During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes …. Check the official notice for the remedy. Verify recall #Z-0219-2024 with the FDA Devices before acting.
The recall
Magellan Diagnostics, Inc. issued this moderate-severity FDA Devices recall-During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0219-2024 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0219-2024) was formally reported on November 15, 2023, with the manufacturer initiating the action on September 19, 2023. It is classified under Moderate severity (Class II), with a current status of Ongoing. Magellan Diagnostics, Inc. is listed as the recalling firm, operating out of North Billerica, MA. Federal records list the affected scope as 11,442 LeadCare II kits (549,216 tests).
The documented reason for this recall is: During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manu… Distribution data in the federal record shows the product reached: US Nationwide distribution.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
11,442 LeadCare II kits (549,216 tests)
Related Recalls
6
3 from same agency
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Ongoing |
| Recall number | Z-0219-2024 |
| Date reported | November 15, 2023 |
| Date initiated | September 19, 2023 |
| Recalling firm | Magellan Diagnostics, Inc. |
| Firm location | North Billerica, MA |
| Affected scope | 11,442 LeadCare II kits (549,216 tests) |
| Distribution | US Nationwide distribution. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 15, 2023. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.