LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.
Product Description
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.
Reason for Recall
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
Details
- Recalling Firm
- Magellan Diagnostics, Inc.
- Units Affected
- 11,442 LeadCare II kits (549,216 tests)
- Distribution
- US Nationwide distribution.
- Location
- North Billerica, MA
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11