LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.

Reported: November 15, 2023 Initiated: September 19, 2023 #Z-0219-2024

Product Description

LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.

Reason for Recall

During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).

Details

Units Affected
11,442 LeadCare II kits (549,216 tests)
Distribution
US Nationwide distribution.
Location
North Billerica, MA

Frequently Asked Questions

What product was recalled?
LeadCare II Blood Lead Test Kit-For in vitro diagnostic use only. The test kit is for the quantitative measurement of lead in capillary whole blood.. Recalled by Magellan Diagnostics, Inc.. Units affected: 11,442 LeadCare II kits (549,216 tests).
Why was this product recalled?
During an internal audit a small percentage (<10%) of plastic caps used to close the treatment reagent tubes were found to not meet the minimum dimensional lip requirement and may have inadequate cap-to-tube seals. Manufacturing with these caps may result in cap-to-tube seals that allow leaking of HCl or may result in cap-to-tube seals that allow evaporation of treatment reagent. The consequence of this would be a change in the volume of treatment reagent and a change in the ratios of either added patient sample or control material to treatment reagent. The result of this might be a falsely elevated result (e.g., a patient false positive [PATFP] result or a Controls Out of Range-High [COOR-Hi] result).
Which agency issued this recall?
This recall was issued by the FDA Devices on November 15, 2023. Severity: Moderate. Recall number: Z-0219-2024.