AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Electrode

Reported: October 30, 2024 Initiated: September 19, 2024 #Z-0221-2025

Product Description

AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Electrode

Reason for Recall

Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.

Details

Units Affected
14 units
Distribution
US Nationwide distribution in the states of CA, OH.
Location
Oak Creek, WI

Frequently Asked Questions

What product was recalled?
AD-TECH Spencer Probe Depth Electrode, REF SD08R-AP58X-000 8-Contact 1.12mm (A style) Depth Electrode. Recalled by Ad-Tech Medical Instrument Corporation. Units affected: 14 units.
Why was this product recalled?
Ad-Tech is recalling these A-Style SD Depth Electrodes because a label discrepancy was identified on A style contact SD electrodes where the MR Conditional symbol was incorrectly placed instead of the MR Unsafe symbol. The A style contacts were not cleared for MR Conditional use. Our records indicate you have received one or more of the affected lots listed above.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 30, 2024. Severity: Moderate. Recall number: Z-0221-2025.