FDA Devices Moderate Class II Terminated

Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal cannula, and a tapered-tip inner needle

Reported: November 24, 2021 Initiated: October 8, 2021 #Z-0223-2022

Product Description

Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal cannula, and a tapered-tip inner needle

Reason for Recall

Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.

Details

Recalling Firm
Cook Inc.
Units Affected
4789 (OUS only)
Distribution
Domestic distribution nationwide. Foreign distribution worldwide.
Location
Bloomington, IN

Frequently Asked Questions

What product was recalled?
Transseptal Needle with Catheter. RPN TSN-17-75.0-ENDRYS. GPN G19261. Coaxial set consisting of an outer catheter, curved-tip metal cannula, and a tapered-tip inner needle. Recalled by Cook Inc.. Units affected: 4789 (OUS only).
Why was this product recalled?
Transseptal needles may contain rust on the interior and/or exterior of the needle, which may not be detected by the user. Potential adverse events if an affected product is used ranges from increased procedural time to inflammatory reactions, which at their most severe include systemic reactions potentially leading to permanent impairment or death.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 24, 2021. Severity: Moderate. Recall number: Z-0223-2022.