PlainRecalls
FDA Devices Moderate Class II Ongoing

Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only

Reported: October 29, 2025 Initiated: September 12, 2025 #Z-0223-2026

Product Description

Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only

Reason for Recall

Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.

Details

Recalling Firm
UIH Technologies LLC
Units Affected
95 units
Distribution
US: OUS:
Location
Houston, TX

Frequently Asked Questions

What product was recalled?
Shanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only. Recalled by UIH Technologies LLC. Units affected: 95 units.
Why was this product recalled?
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
Which agency issued this recall?
This recall was issued by the FDA Devices on October 29, 2025. Severity: Moderate. Recall number: Z-0223-2026.