Amphirion Deep OTW PTA Balloon Catheter: Model numbers: US: AMP015020152 AMP020150152 AMP025120152 AMP030080152 AMP040040152 AMP225210152 AMP020040152 AMP025040152 AMP025150152 AMP030120152 AMP040080152 AMP253210152 AMP020080152 AMP025080002 AMP030040002 AMP030150152 AMP040120152 AMP335210152 AMP020120152 AMP025080152 AMP030040152 AMP035120152 AMP040150152 AMP354210152; OUS: AMD025150152 AMD030150152 AMD040120002 AMD335210002 AMD030040002 AMD035040152 AMD040120152 AMD335210152 AMD03004

Reported: November 18, 2015 Initiated: October 23, 2015 #Z-0243-2016

Product Description

Amphirion Deep OTW PTA Balloon Catheter: Model numbers: US: AMP015020152 AMP020150152 AMP025120152 AMP030080152 AMP040040152 AMP225210152 AMP020040152 AMP025040152 AMP025150152 AMP030120152 AMP040080152 AMP253210152 AMP020080152 AMP025080002 AMP030040002 AMP030150152 AMP040120152 AMP335210152 AMP020120152 AMP025080152 AMP030040152 AMP035120152 AMP040150152 AMP354210152; OUS: AMD025150152 AMD030150152 AMD040120002 AMD335210002 AMD030040002 AMD035040152 AMD040120152 AMD335210152 AMD030040152 AMD035120002 AMD040150002 AMD354210002 AMD030080002 AMD035150002 AMD040150152 AMD354210152 AMD030080152 AMD035150152 AMD225210002 AMD030120002 AMD040040152 AMD225210152 AMD030120152 AMD040080002 AMD253210002 AMD030150002 AMD040080152 AMD253210152 Cardiovascular: Intended for use in Angioplasty

Reason for Recall

Medtronic has discovered that select lots of the Amphirion Deep PTA Balloon Dilation Catheter have a potential breach in the pouch seal enclosing the sterilized product.

Details

Recalling Firm
Medtronic Vascular, Inc.
Units Affected
1652 units in US; 4243 units OUS.
Distribution
Worldwide Distribution-US (nationwide) including Puerto Rico and the countries of Australia Austria Belgium Brazil China Colombia Costa Rica Czech Republic Denmark Dominican Republic Egypt France Georgia Germany Hong Kong Hungary India Iran Ireland Israel Italy Japan Jordan Kazakhstan Korea Kuwait Latvia Lebanon Lithuania Luxembourg Mexico Netherlands Norway Oman Pakistan Panama Peru Philippines Poland Portugal Russian Federation Saudi Arabia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sweden Switzerland Syria Taiwan Thailand Trinidad and Tobago Turkey United Arab Emirates United Kingdom and Uruguay. .
Location
Santa Rosa, CA

Frequently Asked Questions

What product was recalled?
Amphirion Deep OTW PTA Balloon Catheter: Model numbers: US: AMP015020152 AMP020150152 AMP025120152 AMP030080152 AMP040040152 AMP225210152 AMP020040152 AMP025040152 AMP025150152 AMP030120152 AMP040080152 AMP253210152 AMP020080152 AMP025080002 AMP030040002 AMP030150152 AMP040120152 AMP335210152 AMP020120152 AMP025080152 AMP030040152 AMP035120152 AMP040150152 AMP354210152; OUS: AMD025150152 AMD030150152 AMD040120002 AMD335210002 AMD030040002 AMD035040152 AMD040120152 AMD335210152 AMD030040152 AMD035120002 AMD040150002 AMD354210002 AMD030080002 AMD035150002 AMD040150152 AMD354210152 AMD030080152 AMD035150152 AMD225210002 AMD030120002 AMD040040152 AMD225210152 AMD030120152 AMD040080002 AMD253210002 AMD030150002 AMD040080152 AMD253210152 Cardiovascular: Intended for use in Angioplasty. Recalled by Medtronic Vascular, Inc.. Units affected: 1652 units in US; 4243 units OUS..
Why was this product recalled?
Medtronic has discovered that select lots of the Amphirion Deep PTA Balloon Dilation Catheter have a potential breach in the pouch seal enclosing the sterilized product.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 18, 2015. Severity: Moderate. Recall number: Z-0243-2016.