PlainRecalls
ModerateClass IITerminated

FDA Devices recall · Reported November 18, 2015

Cook Medical CXI Support Catheter 2.6FR curved catheter and 4.0FR straight catheter, for cardiovascular use.

Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation re…

Recall #
Z-0244-2016
Affected scope
38
Initiated
October 21, 2015
Verify with FDA Devices →
View my saved recalls

Cook Inc. recalled Cook Medical CXI Support Catheter 2.6FR curved catheter and 4.0FR straight catheter, for… - a moderate-severity action.

Cook Medical CXI Support Catheter 2.6FR curved catheter and 4.0FR straight catheter, for… was recalled by Cook Inc. in November 18, 2015. Reason: Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being m…. Check the official notice for the remedy. Verify recall #Z-0244-2016 with the FDA Devices before acting.

The recall

Cook Inc. issued this moderate-severity FDA Devices recall-Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being m….

Moderate
severity level
38 units
affected scope
Class II
classification
November 18, 2015
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0244-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Cook Medical CXI Support Catheter 2.6FR curved catheter and 4.0FR straight catheter, for cardiovascular use.

Reason for recall

Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation revealed the products were mixed during processing.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Terminated
Recall number Z-0244-2016
Date reported November 18, 2015
Date initiated October 21, 2015
Recalling firm Cook Inc.
Firm location Bloomington, IN
Affected scope 38
Distribution Worldwide Distribution - US Distribution to the states of : CA, FL, ID, IL, MA, NY, OK, TX, UT and VA., and to the countries of Canada, Austria, Germany, Spain, France, Hungary, Sweden and UK.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

Scale of Impact

38 units affected - limited or regional distribution scale.

Regional (<10K units) ✓ This recall
Multi-state (10K – 100K units) -
Large-scale (100K – 1M units) -
Massive (≥1M units) -

Bracket cutoffs follow federal recall-disclosure conventions; bar widths scale linearly within each bracket. Source: PlainRecalls analysis of U.S. Food and Drug Administration filings.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0244-2016) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Cook Medical CXI Support Catheter 2.6FR curved catheter and 4.0FR straight catheter, for cardiovascular use.. Recalled by Cook Inc.. Units affected: 38.
Why was this product recalled?
Cook Medical has received six product complaints associated with reports of the 2.6FR curved catheter being mixed with the 4.0FR straight catheter and vice versa. Investigation revealed the products were mixed during processing.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 18, 2015. Severity: Moderate. Recall number: Z-0244-2016.
Where was the recalled product distributed?
Distribution: Worldwide Distribution - US Distribution to the states of : CA, FL, ID, IL, MA, NY, OK, TX, UT and VA., and to the countries of Canada, Austria, Germany, Spain, France, Hungary, Sweden and UK..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported November 18, 2015. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026