Severity
Moderate
FDA Devices recall · Reported November 18, 2015
Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) f…
Maquet Cardiovascular Us Sales, Llc recalled Maquet Getinge Group, Adjustable Positioning Elements (1007.28A0) Manufacturer: Maquet Gm… - a moderate-severity action.
Maquet Getinge Group, Adjustable Positioning Elements (1007.28A0) Manufacturer: Maquet Gm… was recalled by Maquet Cardiovascular Us Sales, Llc in November 18, 2015. Reason: Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 10…. Check the official notice for the remedy. Verify recall #Z-0247-2016 with the FDA Devices before acting.
The recall
Maquet Cardiovascular Us Sales, Llc issued this moderate-severity FDA Devices recall-Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 10….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0247-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0247-2016) was formally reported on November 18, 2015, with the manufacturer initiating the action on August 17, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Maquet Cardiovascular Us Sales, Llc is listed as the recalling firm, operating out of Wayne, NJ. Federal records list the affected scope as United States (3 sets); Rest of World (47 sets).
The documented reason for this recall is: Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elemen… Distribution data in the federal record shows the product reached: Worldwide Distribution - US including AL, CA, MO and Internationally to Australia, Brazil, Canada, China, Croatia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Russia, Singapore, South Africa and …. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
United States (3 sets); Rest of World (47 sets)
Related Recalls
6
3 from same agency
Maquet Getinge Group, Adjustable Positioning Elements (1007.28A0) Manufacturer: Maquet GmbH, Germany. Patient positioners. The MAQUET Adjustable Positioning Elements are designed for the placement and positioning of patients in the prone position immediately before, during and after surgical interventions as well as for examination and treatment. With a patient in the prone position, the adjustable positioning elements stabilize the upper body while simultaneously relieving the abdominal region in the vertical position with regard to the spine and the large blood vessels there. This positioning method is thus suitable for interventions on the spine with dorsal access to prevent positioning-specific compression with resulting bleeding tenancies. They are made up of the following components: (2) chest and (2) hip elements with padding which are secured in a fixed arrangement to the frames and can be adjusted in all directions independently of each other.
Two serious thromboembolic events were reported involving the Adjustable Positioning Elements (Part Number 1007.28A0) in conjunction with Universal Frame (Part Number 1007.24F0) for lumbar spine surgery. The hip elements may have a negative effect on patients- venous backflow under certain conditions and may have been a contributing factor to thromboembolic events.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0247-2016 |
| Date reported | November 18, 2015 |
| Date initiated | August 17, 2015 |
| Recalling firm | Maquet Cardiovascular Us Sales, Llc |
| Firm location | Wayne, NJ |
| Affected scope | United States (3 sets); Rest of World (47 sets) |
| Distribution | Worldwide Distribution - US including AL, CA, MO and Internationally to Australia, Brazil, Canada, China, Croatia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Russia, Singapore, South Africa and United Arab Emirates. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 18, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.