R-20J , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-Ray tubes: 06/1.2P364DK-85, 0.6/1.2P324DK-125, 06/1.2P324DK-85, 0.6/1.2P38DE-85. Product Usage: The device is the beam-limiting device for X-ray Radiography.

Reported: November 14, 2012 Initiated: October 28, 2011 #Z-0261-2013

Product Description

R-20J , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-Ray tubes: 06/1.2P364DK-85, 0.6/1.2P324DK-125, 06/1.2P324DK-85, 0.6/1.2P38DE-85. Product Usage: The device is the beam-limiting device for X-ray Radiography.

Reason for Recall

Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.

Details

Recalling Firm
Shimadzu Medical Systems
Units Affected
308 units
Distribution
US Nationwide Distribution
Location
Torrance, CA

Frequently Asked Questions

What product was recalled?
R-20J , Diagnostic X-Ray Beam-Limiting Device. In combination with any of below X-Ray tubes: 06/1.2P364DK-85, 0.6/1.2P324DK-125, 06/1.2P324DK-85, 0.6/1.2P38DE-85. Product Usage: The device is the beam-limiting device for X-ray Radiography.. Recalled by Shimadzu Medical Systems. Units affected: 308 units.
Why was this product recalled?
Shimadzu is recalling Diagnostic X-ray Beam-Limiting Device, Mobile x-ray system and Diagnostic x-ray tube assembly because they do not comply with the Product Performance Standards, 21CFR 1020.30 (m) Beam quality --(1) Half-value layer (HVL), which was revised in 2006. Certain devices might not comply with the requirements for them in some x-ray generating condition after the revision in 2006.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 14, 2012. Severity: Moderate. Recall number: Z-0261-2013.