Severity
Moderate
FDA Devices recall · Reported November 20, 2013
On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522-000 Rev. H). On March 15, 2013, when the product was going to be pla…
Stryker Instruments Div. of Stryker Corporation recalled Acetabular Pressurizer; single use and six units per box. The Acetabular Pressurizer 0… - a moderate-severity action.
Acetabular Pressurizer; single use and six units per box. The Acetabular Pressurizer 0… was recalled by Stryker Instruments Div. of Stryker Corporation in November 20, 2013. Reason: On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-52…. Check the official notice for the remedy. Verify recall #Z-0280-2014 with the FDA Devices before acting.
The recall
Stryker Instruments Div. of Stryker Corporation issued this moderate-severity FDA Devices recall-On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-52….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0280-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0280-2014) was formally reported on November 20, 2013, with the manufacturer initiating the action on November 1, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Stryker Instruments Div. of Stryker Corporation is listed as the recalling firm, operating out of Portage, MI. Federal records list the affected scope as 87 boxes (522 each).
The documented reason for this recall is: On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522-000 Rev. H). On March 15, 2013, when the product was going to be placed in the pouch to be sealed, the manuf… Distribution data in the federal record shows the product reached: Worldwide distribution: US (nationwide) including states of: TX, FL, IL, IA, ND, WI, VA, and NJ.; and countries of: CANADA, POLAND and SWITZERLAND.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
87 boxes (522 each)
Related Recalls
6
3 from same agency
Acetabular Pressurizer; single use and six units per box. The Acetabular Pressurizer 0206-522-000 is an accessory to the Advanced Cement Mixer (ACM). The Acetabular Pressurizer Tiplet P/N 0206-522-012 is sealed in a two pouches, sterilized, and labeled as part number 0206-522-000. This Acetabular Pressurizer is an accessory used to pressurize the cement applied to the prosthesis to be implanted during the surgery. The Acetabular Pressurizer is an injection molded component made with resin material DOW HDPD 10462N which is a rigid plastic once it is molded.
On March 14, 2013 a Work Order was opened for 120 units (20 boxes) of the Acetabular Pressurizer (P/N 0206-522-000 Rev. H). On March 15, 2013, when the product was going to be placed in the pouch to be sealed, the manufacturing operator noticed that some assemblies were cracked. As a result of the injection pressure of the cement gun, a cracked tiplet may release from the Hub during cement applica
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0280-2014 |
| Date reported | November 20, 2013 |
| Date initiated | November 1, 2013 |
| Recalling firm | Stryker Instruments Div. of Stryker Corporation |
| Firm location | Portage, MI |
| Affected scope | 87 boxes (522 each) |
| Distribution | Worldwide distribution: US (nationwide) including states of: TX, FL, IL, IA, ND, WI, VA, and NJ.; and countries of: CANADA, POLAND and SWITZERLAND. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 20, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.