Reported: November 15, 2023 Initiated: October 12, 2023 #Z-0290-2024

Product Description

Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder)

Reason for Recall

It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.

Details

Units Affected
1022 units
Distribution
US Nationwide.
Location
Wayne, NJ

Frequently Asked Questions

What product was recalled?
Cardiohelp-i, blood oxygenation and carbon dioxide removal system, Part Numbers 701072780 (Cardiohelp-i), 701068550 (Slide Rail Holder). Recalled by Maquet Medical Systems USA. Units affected: 1022 units.
Why was this product recalled?
It was found that the fixation (weld) employed by the supplier of the nut bolt to the slide rail plate was insufficient. The deficient weld allows the nut bolt to break off from the slide rail plate with slight manual force.