Medtronic Wireless Recharger Model WR9200 included in the RS6200 Recharger Kit used by the Activa RC (Model 37612) implantable neurostimulator.
Reported: December 1, 2021 Initiated: November 4, 2021 #Z-0294-2022
Product Description
Medtronic Wireless Recharger Model WR9200 included in the RS6200 Recharger Kit used by the Activa RC (Model 37612) implantable neurostimulator.
Reason for Recall
The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
Details
- Recalling Firm
- Medtronic Neuromodulation
- Units Affected
- 6,406 devices
- Distribution
- US Nationwide, including PR. There was government distribution but no military distribution was identified.
- Location
- Minneapolis, MN
Frequently Asked Questions
What product was recalled? ▼
Medtronic Wireless Recharger Model WR9200 included in the RS6200 Recharger Kit used by the Activa RC (Model 37612) implantable neurostimulator.. Recalled by Medtronic Neuromodulation. Units affected: 6,406 devices.
Why was this product recalled? ▼
The wireless rechargers (WR; WR9200 and WR9220; included in the Recharger Kits) become unresponsive if the charging steps, indicated in the Recharger Patient User Guide, are not followed.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on December 1, 2021. Severity: Moderate. Recall number: Z-0294-2022.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11