4205 Journey Rolling Walker Product Usage: used to assist people in walking
Reported: November 2, 2016 Initiated: October 1, 2016 #Z-0304-2017
Product Description
4205 Journey Rolling Walker Product Usage: used to assist people in walking
Reason for Recall
Nova Medical is recalling the Journey Rolling Walker because the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out.
Details
- Recalling Firm
- Nova Ortho-Med Inc
- Units Affected
- 1,380 units
- Distribution
- US Nationwide Distribution in the states of: CA, IN, HI, CT, UT, WA, FL, VA, WI, OH, MN, KS, MD, IL, MO, WY, MT, ME, PA, TN, MI, OR, NY, NJ, ID, NC, IA, OK, AR, VT, SC, TX
- Location
- Carson, CA
Frequently Asked Questions
What product was recalled? ▼
4205 Journey Rolling Walker Product Usage: used to assist people in walking. Recalled by Nova Ortho-Med Inc. Units affected: 1,380 units.
Why was this product recalled? ▼
Nova Medical is recalling the Journey Rolling Walker because the screws used to hold the black side support pieces to the frame can become loose over time. This can result in the locking nut coming off and the screw falling out.
Which agency issued this recall? ▼
This recall was issued by the FDA Devices on November 2, 2016. Severity: Moderate. Recall number: Z-0304-2017.
Related Recalls
FDA Devices Moderate
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
FDA Devices Moderate
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
FDA Devices Moderate
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
FDA Devices Moderate
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11