BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364, 44336409 (Repaired Reader spare part) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Reported: November 7, 2018 Initiated: September 24, 2018 #Z-0312-2019

Product Description

BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364, 44336409 (Repaired Reader spare part) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.

Reason for Recall

A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive.

Details

Recalling Firm
Becton Dickinson & Co.
Units Affected
23
Distribution
Worldwide distribution - US nationwide in the states of AL, AZ, CA, CO, DC, FL, GA, IL, IN, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, OR, PA, SC, TN, TX, UT, VA, WA, WI, and Puerto Rico. Countries of Spain, Germany, France, Austria, Denmark, Switzerland, Sweden, Italy, Kuwait, Oman, Saudi Arabia, and Turkey.
Location
Sparks, MD

Frequently Asked Questions

What product was recalled?
BD Max Instrument Reader Spare Part, Catalog Number 8090369/443364, 44336409 (Repaired Reader spare part) - Product Usage: The BD MAX System, used in conjunction with appropriate extraction kits and reagents, is capable of automated extraction and purification of nucleic acids from multiple specimen types. The BD MAX System is also capable of automated amplification and detection of target nucleic acid sequences for use in clinical, industrial, and research settings.. Recalled by Becton Dickinson & Co.. Units affected: 23.
Why was this product recalled?
A trend for false positive Candida glabrata results has been confirmed. The discrepant results impact the Candida glabrata target only, when in the presence of a Trichomonas vaginalis true positive.
Which agency issued this recall?
This recall was issued by the FDA Devices on November 7, 2018. Severity: Moderate. Recall number: Z-0312-2019.