Severity
Moderate
FDA Devices recall · Reported November 27, 2013
There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an unlocked position this may, under certain circumstances, cause the pum…
Olympus America Inc. recalled Olympus Endoscopic Flushing Pump, Model OPF-2. Facilitates washing of tissues to remov… - a moderate-severity action.
Olympus Endoscopic Flushing Pump, Model OPF-2. Facilitates washing of tissues to remov… was recalled by Olympus America Inc. in November 27, 2013. Reason: There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an …. Check the official notice for the remedy. Verify recall #Z-0329-2014 with the FDA Devices before acting.
The recall
Olympus America Inc. issued this moderate-severity FDA Devices recall-There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an ….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0329-2014 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0329-2014) was formally reported on November 27, 2013, with the manufacturer initiating the action on May 20, 2013. It is classified under Moderate severity (Class II), with a current status of Terminated. Olympus America Inc. is listed as the recalling firm, operating out of Center Valley, PA. Federal records list the affected scope as 263.
The documented reason for this recall is: There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an unlocked position this may, under certain circumstances, cause the pump to stop and go into power stand-by mod… Distribution data in the federal record shows the product reached: Worldwide distribution: USA (nationwide) and the country of Canada.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
263
Related Recalls
6
3 from same agency
Olympus Endoscopic Flushing Pump, Model OPF-2. Facilitates washing of tissues to remove blood, feces and other organic matter during endoscopic procedures.
There is a possibility that the pump head was not shipped in the locked position. If the pump head is in an unlocked position this may, under certain circumstances, cause the pump to stop and go into power stand-by mode.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0329-2014 |
| Date reported | November 27, 2013 |
| Date initiated | May 20, 2013 |
| Recalling firm | Olympus America Inc. |
| Firm location | Center Valley, PA |
| Affected scope | 263 |
| Distribution | Worldwide distribution: USA (nationwide) and the country of Canada. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 27, 2013. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.