PlainRecalls
CriticalClass IOngoing

FDA Devices recall · Reported December 22, 2021

Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.

Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which c…

Recall #
Z-0336-2022
Affected scope
5,566 (US); 6,354 (OUS)
Initiated
November 15, 2021
Verify with FDA Devices →
View my saved recalls

Datascope Corp. recalled Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-… - a critical-severity action.

Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-… was recalled by Datascope Corp. in December 22, 2021. Reason: Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short …. Check the official notice for the remedy. Verify recall #Z-0336-2022 with the FDA Devices before acting.

The recall

Datascope Corp. issued this critical-severity FDA Devices recall-Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short ….

Critical
severity level
Class I
classification
December 22, 2021
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0336-2022 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This FDA Devices action (record #Z-0336-2022) was formally reported on December 22, 2021, with the manufacturer initiating the action on November 15, 2021. It is classified under Critical severity (Class I), with a current status of Ongoing. Datascope Corp. is listed as the recalling firm, operating out of Mahwah, NJ. Federal records list the affected scope as 5,566 (US); 6,354 (OUS).

The documented reason for this recall is: Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instab… Distribution data in the federal record shows the product reached: Worldwide distribution - US Nationwide.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 40,861 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,787
Severity1902064468High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,001 recalls with an agency hazard classification, 19,020 are high severity, 64,468 moderate, and 4,513 low. A further 16,157 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified high severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.

Reason for recall

Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Critical (Class I)
Status Ongoing
Recall number Z-0336-2022
Date reported December 22, 2021
Date initiated November 15, 2021
Recalling firm Datascope Corp.
Firm location Mahwah, NJ
Affected scope 5,566 (US); 6,354 (OUS)
Distribution Worldwide distribution - US Nationwide.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0336-2022) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • This is a Class I (high-risk) recall, stop using the product immediately and follow the disposal or return instructions. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Cardiosave Hybrid IntraAortic Balloon Pump. Model Numbers: 0998-00-0800-31, 0998-UC-0800-31, 0998-00-0800-32, 0998-UC-0800-32, 0998-00-0800-33, 0998-UC-0800-33, 0998-00-0800-34, 0998-UC-0800-34, 0998-00-0800-35, 0998-UC-0800-35, 0998-00-0800-36, 0998-UC-0800-36, 0998-00-0800-45, 0998-UC-0800-45, 0998-00-0800-52, 0998-UC-0800-52, 0998-00-0800-53, 0998-UC-0800-53, 0998-00-0800-55, 0998-UC-0800-55, 0998-00-0800-65, 0998-UC-0800-65.. Recalled by Datascope Corp.. Units affected: 5,566 (US); 6,354 (OUS).
Why was this product recalled?
Firm is initiating a correction due to the possibility of fluid ingress. Fluid entering the device may short various electronic components thus leading to system shutdown, which could result in unsafe hemodynamic instability.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 22, 2021. Severity: Critical. Recall number: Z-0336-2022.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Similar Medical Devices · nearest report date

Same product category, closest by report date among other firms and states (deduped against the severity set).

Compare this recall with FARADRIVE STEERABLE SHEATH CLEAR, used to facilitate access… → · All Medical Devices recalls

This recall: FDA Devices, reported December 22, 2021. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026