GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Analyte Specific Reagent, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.
Product Description
GREER M. FAENI, 0.25 mL 20 MG/ML, Item: K6, Analyte Specific Reagent, HYPERSENSITIVITY PNEUMONITIS test system, gel diffusion kit for demonstrating precipitating antibodies.
Reason for Recall
The product was recalled because stability evaluation of the lots indicated a decrease in reactivity that could lead to a potential false-negative result interpretation. The issue was identified by the firm after failing to meet lot requirements for potency. The lot release error was discovered on October 08, 2021, after a failed potency test was determined during 6-month stability evaluation of retention material for the lots. The subsequent manufacturing investigation, which included a review of the lot release potency documentation, found that the initial lot release potency test for the lot was misinterpreted.
Details
- Recalling Firm
- Greer Laboratories, Inc.
- Units Affected
- 218 units
- Distribution
- US Distribution to states of: CA, CO, IA, KY, MA, MD, MO, NC, PA, UT, VA, and International to countries of: Canada, Denmark
- Location
- Lenoir, NC
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11