LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause th…
Smith & Nephew, Inc. recalled LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails… - a moderate-severity action.
LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails… was recalled by Smith & Nephew, Inc. in November 20, 2019. Reason: LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide…. Check the official notice for the remedy. Verify recall #Z-0351-2020 with the FDA Devices before acting.
The recall
Smith & Nephew, Inc. issued this moderate-severity FDA Devices recall-LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0351-2020 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0351-2020) was formally reported on November 20, 2019, with the manufacturer initiating the action on October 7, 2019. It is classified under Moderate severity (Class II), with a current status of Terminated. Smith & Nephew, Inc. is listed as the recalling firm, operating out of Memphis, TN. Federal records list the affected scope as 116 devices.
The documented reason for this recall is: LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress. Distribution data in the federal record shows the product reached: Worldwide distributions - US Nationwide distributions and countries of Belgium, Brazil, China, Columbia, France, Great Britain, India, Japan, Russia, Singapore, South Africa, Spain,and United Arab Emirates.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
116 devices
Related Recalls
6
3 from same agency
LAG SCREW 3.2MM GUIDE PIN SLEEVE, REF 71674032 - Product Usage: The TRIGEN InterTAN nails are indicated for fractures of the femur including: simple shaft fractures, comminuted shaft fractures, spiral shaft fractures, long oblique shaft fractures and segmental shaft fractures; subtrochanteric fractures; intertrochanteric fractures; ipsilateral femoral shaft/neck fractures; intracapsular fractures; nonunions and malunions; polytrauma and multiple fractures; prophylactic nailing of impending pathologic fractures; reconstruction, following tumor resection and grafting; bone lengthening and shortening.
LAG SCREW 3.2 GUIDE PIN SLEEVE are being recalled due to a manufacturing error. The detent balls in the guide pin sleeve were not drilled to the correct depth, which will cause them to not fully compress.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0351-2020 |
| Date reported | November 20, 2019 |
| Date initiated | October 7, 2019 |
| Recalling firm | Smith & Nephew, Inc. |
| Firm location | Memphis, TN |
| Affected scope | 116 devices |
| Distribution | Worldwide distributions - US Nationwide distributions and countries of Belgium, Brazil, China, Columbia, France, Great Britain, India, Japan, Russia, Singapore, South Africa, Spain,and United Arab Emirates. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported November 20, 2019. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.