Severity
Moderate
FDA Devices recall · Reported December 9, 2015
Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some…
ConMed Corporation recalled Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500… - a moderate-severity action.
Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500… was recalled by ConMed Corporation in December 9, 2015. Reason: Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperatur…. Check the official notice for the remedy. Verify recall #Z-0359-2016 with the FDA Devices before acting.
The recall
ConMed Corporation issued this moderate-severity FDA Devices recall-Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperatur….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0359-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0359-2016) was formally reported on December 9, 2015, with the manufacturer initiating the action on November 5, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. ConMed Corporation is listed as the recalling firm, operating out of Utica, NY. Federal records list the affected scope as 9 devices.
The documented reason for this recall is: Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" mess… Distribution data in the federal record shows the product reached: Distributed in the states of CA, and TX and in the country of Mexico.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
9 devices
Related Recalls
6
3 from same agency
Altrus Thermal Tissue Fusion Energy Sources, Catalog Numbers: 1) 60-9500-120 & 2) 60-9500-INT. Thermal Ligation and Cutting Energy Source.
Miscalibration of the Energy Source monitoring function, which may incorrectly sense the handpiece temperature as higher than the actual handpiece's temperature. This causes some devices to generate a "Check Seal" message with an audible alarm and to stop the heating cycle.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0359-2016 |
| Date reported | December 9, 2015 |
| Date initiated | November 5, 2015 |
| Recalling firm | ConMed Corporation |
| Firm location | Utica, NY |
| Affected scope | 9 devices |
| Distribution | Distributed in the states of CA, and TX and in the country of Mexico. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 9, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.