PlainRecalls
ModerateClass IIOngoing

FDA Devices recall · Reported November 20, 2024

Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-046-001

There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the…

Recall #
Z-0369-2025
Affected scope
21,490 eaches
Initiated
October 4, 2024
Verify with FDA Devices →
View my saved recalls

Stryker Corporation recalled Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Desi… - a moderate-severity action.

Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Desi… was recalled by Stryker Corporation in November 20, 2024. Reason: There is a potential for the product to activate without manual input when plugged into a power source or rem…. Check the official notice for the remedy. Verify recall #Z-0369-2025 with the FDA Devices before acting.

The recall

Stryker Corporation issued this moderate-severity FDA Devices recall-There is a potential for the product to activate without manual input when plugged into a power source or rem….

Moderate
severity level
Class II
classification
November 20, 2024
reported

Sourced from official FDA Devices enforcement records. Verify recall #Z-0369-2025 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.

Recall Insight

This recall's affected scope, documented in the Recall Profile below, stays within a single-region or limited-distribution footprint. Distribution scope (see Distribution below) directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.

Within the same product category the archive holds 6 closely related recalls, 6 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.

Medical Devices recalls over time

Where this recall sits in its category - 41,021 medical devices recalls on record

01,0002,0003,0004,000 20052008201120142017202020232026 1,947
Severity1906864652High severity (most serious)Moderate severityLow severity
Where this agency-classified recall sits among FDA Class I/II/III notices

Of 88,244 recalls with an agency hazard classification, 19,068 are high severity, 64,652 moderate, and 4,524 low. A further 16,179 CPSC/NHTSA recalls have no agency class and are excluded from this split. This recall is classified moderate severity.

Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.

Product description

Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-046-001

Reason for recall

There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries

Recall Profile

Structured summary of the FDA Devices recall record
Attribute Value
Agency U.S. Food and Drug Administration
Severity class Moderate (Class II)
Status Ongoing
Recall number Z-0369-2025
Date reported November 20, 2024
Date initiated October 4, 2024
Recalling firm Stryker Corporation
Firm location Portage, MI
Affected scope 21,490 eaches
Distribution Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia.

Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.

What to do with this recall

Match your product against the recall record, then act on the agency remedy.

  • Check the recall number (Z-0369-2025) and product description against the item you own. Search the archive
  • Confirm the current status with FDA Devices before acting, recall details can be updated.
  • Follow the documented remedy (refund, replacement, repair, or disposal) rather than general advice. What to do next

This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.

Frequently Asked Questions

What product was recalled?
Stryker Neptune SafeAir Smoke Evacuation Pencil Non coated 70 mm Blade Push-Button - Designed for general electrosurgical applications including cutting and coagulation and for removing smoke generated by electrosurgery when used in conjunction with an effective smoke evacuation system. Catalog Number: 0703-046-001. Recalled by Stryker Corporation. Units affected: 21,490 eaches.
Why was this product recalled?
There is a potential for the product to activate without manual input when plugged into a power source or remain active after buttons are released. Unintentional activation of the device may result in localized burns to the patient or physician from electrical currents carried through conductive objects or fluids. Medical/surgical intervention may be needed to correct these injuries
Which agency issued this recall?
This recall was issued by the FDA Devices on November 20, 2024. Severity: Moderate. Recall number: Z-0369-2025.
Where was the recalled product distributed?
Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Japan, Australia..
Should I stop using a recalled medication or medical device?
Do not stop using a recalled medication or device without consulting your healthcare provider first, as abruptly discontinuing treatment could pose its own health risks. Your doctor can advise on alternatives or whether the recall applies to your specific product lot. Check the recall number and lot information against your product packaging.

Nationwide recalls with similar severity or category timing

Drug and device notices outside this firm and state, matched by severity timing and product category.

Compare this recall with GORE VIABIL Short Wire Biliary Endoprosthesis, Model No. VS… → · All Medical Devices recalls

This recall: FDA Devices, reported November 20, 2024. Severity classes follow each agency's own taxonomy; related-recall context is computed across the full archive.

Methodology · Corrections · Archive as of August 2026