FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE CHINA

Reported: December 14, 2022 Initiated: August 31, 2022 #Z-0370-2023

Product Description

FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE CHINA

Reason for Recall

This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.

Details

Recalling Firm
Full Vision Inc
Units Affected
8 units
Distribution
Wi, GA, NY, Mexico
Location
Newton, KS

Frequently Asked Questions

What product was recalled?
FULL VISION INC. TRACKMASTER, TREADMILL GE T2100-ST2 220V, Model #317-07927GE CHINA. Recalled by Full Vision Inc. Units affected: 8 units.
Why was this product recalled?
This has been identified a rare occurrence, under a specific scenario, where the circuit is found in a latch-up condition on the PCB that controls the safety function of the Magnetic Safety Tether, that when the magnet is pulled loose as during a fall, the treadmill may fail to stop.
Which agency issued this recall?
This recall was issued by the FDA Devices on December 14, 2022. Severity: Moderate. Recall number: Z-0370-2023.