The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedure in which a refluxing vein is thermally coagulated to permanently eliminate the vein from blood circulation. The catheter has a long shaft for insertion into a vein, a handle with start/stop button, and a connector cable. The catheter shaft has a heating element that is energized by the digiRF radiofrequency generator, which is a multi-voltage energy delivery system with tou
Product Description
The EVSRF catheter is a sterile, single-use disposable medical device for endovenous radiofrequency ablation, a procedure in which a refluxing vein is thermally coagulated to permanently eliminate the vein from blood circulation. The catheter has a long shaft for insertion into a vein, a handle with start/stop button, and a connector cable. The catheter shaft has a heating element that is energized by the digiRF radiofrequency generator, which is a multi-voltage energy delivery system with touchscreen control that automatically sets the non-adjustable treatment parameters for the catheter. A button on the catheter (or a foot pedal attached to the generator) begins an automated treatment cycle of 20 seconds long at a set temperature of 120 ¿C, and the treatment stops automatically when complete. The catheter is used within a sterile operative field, while the generator remains outside the sterile operative field at all times.
Reason for Recall
The Venclose" RF Ablation Catheters may have been manufactured with internal wiring error. The red and yellow signal wires were soldered on swapped ports of the catheter PCB. This error in wire configuration would lead to power being applied through the proximal 3/4th of the heating coil, bypassing the thermocouple, which will then cause the generator to overdrive the catheter in an attempt to reach desired temperature set point. The wire attachment error causes the device to reach temperatures exceeding the levels intended for therapeutic efficacy, while the console is falsely indicating that temperature has not yet been reached, thus increasing the risk of harm due to excessive temperature.
Details
- Recalling Firm
- Bard Peripheral Vascular Inc
- Units Affected
- 222,373 catheters
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI and WV. The countries of Australia, Singapore, Austria, Belgium, Czechia, Denmark, France, Germany, Greece, Ireland, Italy, Poland, Slovenia Spain, Sweden, Switzerland, United Kingdom, United Arab Emirates, and South Africa.
- Location
- Tempe, AZ
Frequently Asked Questions
What product was recalled? ▼
Why was this product recalled? ▼
Which agency issued this recall? ▼
Related Recalls
Convenience Kits 1. 25GA Convenience Kit - FNX VS0644.25-FNX 2. 25GA Convenience Kit - FXX VS0644.25-FXX 3.2…
Vortex Surgical Inc. · 2026-02-11
COULTER DxH Diluent, REF 628017 COULTER DxH ECO Diluent REF C67250 For use as an isotonic buffered diluent in conju…
Beckman Coulter, Inc. · 2026-02-11
Product Name: Single Use 3-Lumen Sphincterotome V Model/Catalog Number: KD-V411M-1520 Product Description: These inst…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Disposable Triple Lumen Sphincterotome Model/Catalog Number: KD-401Q-0330 Product Description: These in…
Olympus Corporation of the Americas · 2026-02-11
Product Name: Lumipulse pTau 217 Plasma Controls Model/Catalog Number: 81558 Software Version: Not Applicable Produc…
Fujirebio Diagnostics, Inc. · 2026-02-11