Severity
Moderate
FDA Devices recall · Reported December 16, 2015
TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the …
Synthes (USA) Products LLC recalled Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile… - a moderate-severity action.
Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile… was recalled by Synthes (USA) Products LLC in December 16, 2015. Reason: TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulat…. Check the official notice for the remedy. Verify recall #Z-0375-2016 with the FDA Devices before acting.
The recall
Synthes (USA) Products LLC issued this moderate-severity FDA Devices recall-TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulat….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0375-2016 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0375-2016) was formally reported on December 16, 2015, with the manufacturer initiating the action on November 2, 2015. It is classified under Moderate severity (Class II), with a current status of Terminated. Synthes (USA) Products LLC is listed as the recalling firm, operating out of West Chester, PA. Federal records list the affected scope as 5.
The documented reason for this recall is: TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling na… Distribution data in the federal record shows the product reached: US Distribution to states of: CA, FL, MI and WI.. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
5
Related Recalls
6
3 from same agency
Trochanteric Fixation Nail- Advanced System-11MM/130 DEG TI CANN TFNA 420MM/Right Sterile (The TFN-ADVANCED Proximal Femoral Nailing System) The TFN-ADVANCED Proximal Femoral Nailing System is intended for treatment of fractures. Intramedullary fixation rod, orthopedic.
TFNA nails may have the locking mechanism too close to the top of the nail. This may 1. Prevent the cannulated connecting screw from fully tightening the insertion handle to the nail resulting in a loose or toggling nail in the handle. Or 2. Cause the bottom of the cannulated connecting screw to tighten against the top of the locking mechanism preventing advancement of the locking mechanism.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0375-2016 |
| Date reported | December 16, 2015 |
| Date initiated | November 2, 2015 |
| Recalling firm | Synthes (USA) Products LLC |
| Firm location | West Chester, PA |
| Affected scope | 5 |
| Distribution | US Distribution to states of: CA, FL, MI and WI. |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported December 16, 2015. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.