Severity
Moderate
FDA Devices recall · Reported January 24, 2018
SPAG-2 unit nebulizers from Lot J0571670RA were tested at Next Breath for particle sizing under the validated method, NB-TM-261, In November 2016. The particle size specification …
Valeant Pharmacueticals International recalled Small Particle Aerosol Generator Model SPAG-2 6000 Series (SPAG-2). The SPAG-2 medical de… - a moderate-severity action.
Small Particle Aerosol Generator Model SPAG-2 6000 Series (SPAG-2). The SPAG-2 medical de… was recalled by Valeant Pharmacueticals International in January 24, 2018. Reason: SPAG-2 unit nebulizers from Lot J0571670RA were tested at Next Breath for particle sizing under the validated…. Check the official notice for the remedy. Verify recall #Z-0384-2018 with the FDA Devices before acting.
The recall
Valeant Pharmacueticals International issued this moderate-severity FDA Devices recall-SPAG-2 unit nebulizers from Lot J0571670RA were tested at Next Breath for particle sizing under the validated….
Sourced from official FDA Devices enforcement records. Verify recall #Z-0384-2018 with the agency before acting. Full product description, hazard, remedy, and related recalls are below.
This FDA Devices action (record #Z-0384-2018) was formally reported on January 24, 2018, with the manufacturer initiating the action on March 13, 2017. It is classified under Moderate severity (Class II), with a current status of Terminated. Valeant Pharmacueticals International is listed as the recalling firm, operating out of Rochester, NY. Federal records list the affected scope as 418 SPAG-2 and 808 Nebulizers.
The documented reason for this recall is: SPAG-2 unit nebulizers from Lot J0571670RA were tested at Next Breath for particle sizing under the validated method, NB-TM-261, In November 2016. The particle size specification set in NB-SPEC-039 for Dv10, Dv50, & Dv9… Distribution data in the federal record shows the product reached: US Nationwide & Canada. Distribution scope directly affects the consumer exposure window and determines whether a recall remains regional or escalates into a nationwide advisory.
Within the same product category the archive holds 6 closely related recalls, 3 from FDA Devices - clustering in a narrow category often points to a systemic quality-control or supplier issue rather than a one-off defect. Always verify the recall number against the official agency record before acting.
Where this recall sits in its category - 40,322 medical devices recalls on record
Of 101,704 recalls in the database, 24,878 are high severity, 72,426 moderate, and 4,400 low. This recall is classified moderate severity.
Counts reflect market size and reporting activity, not inherent danger, we do not rank products by risk from raw recall volume.
Severity
Moderate
Affected scope
418 SPAG-2 and 808 Nebulizers
Related Recalls
6
3 from same agency
Small Particle Aerosol Generator Model SPAG-2 6000 Series (SPAG-2). The SPAG-2 medical device is a pneumatic flow system indicated for the administration of Virazole (ribavirin) aerosol. Driven by a pressurized gas (air or blended oxygen) and uses a nebulizer to generate a fine aerosol of hydrated Virazole. --- CDRH Classification Name: Nebulizer (Direct Patient Interface)
SPAG-2 unit nebulizers from Lot J0571670RA were tested at Next Breath for particle sizing under the validated method, NB-TM-261, In November 2016. The particle size specification set in NB-SPEC-039 for Dv10, Dv50, & Dv90 are a mean value of no more than 3.0¿m, 5.5 ¿m, and 15¿m respectively. Initial testing of the nebulizers yielded four units that delivered particle sizes that did not meet the specification (NB-SPEC-039). The out of specification (OOS) nebulizers were cleaned per the approved SPAG-2 procedure, SPE-000127228 Rev 2, to remove any product residue and were then re-evaluated. Secondary testing yielded that three of the four nebulizers still did not meet the particle size specification. It was concluded that the 3 nebulizers from SPAG-2 Lot J0571670RA failed to meet (SPE-000127228 Rev 2) did not resolve the high droplet size distribution values observed.
| Attribute | Value |
|---|---|
| Agency | U.S. Food and Drug Administration |
| Severity class | Moderate (Class II) |
| Status | Terminated |
| Recall number | Z-0384-2018 |
| Date reported | January 24, 2018 |
| Date initiated | March 13, 2017 |
| Recalling firm | Valeant Pharmacueticals International |
| Firm location | Rochester, NY |
| Affected scope | 418 SPAG-2 and 808 Nebulizers |
| Distribution | US Nationwide & Canada |
Profile values are sourced directly from the official FDA Devices enforcement record. Source: U.S. Food and Drug Administration.
What to do with this recall
Match your product against the recall record, then act on the agency remedy.
This page summarizes the official FDA Devices record for research and awareness; it is not legal, medical, or safety advice. Verify with the issuing agency before acting.
Other recalls in the same product category, useful for spotting patterns across the same defect class or manufacturer.
· 2026-07-02
· 2026-06-25
· 2026-06-18
GE Medical Systems Information Technologies Inc · 2026-06-03
Medline Industries, LP · 2026-06-03
Compare this recall with Rowenta Recalls Cordless Vacuum Cleaners Due to Risk of Ser… →
Every figure on PlainRecalls is rendered directly from official FDA, CPSC and NHTSA recall records, no number is typed in by an editor. This recall: FDA Devices, reported January 24, 2018. Severity classes follow each agency's own taxonomy (FDA Class I/II/III; CPSC and NHTSA by hazard type), and related-recall context is computed across the full archive. See our editorial standards & corrections policy, the methodology behind these numbers, or report a data error. Data current as of July 2026.